SECTION 4 PREPARATION FOR USE AND INSTALLATION
_______________________________________________________________________
4-4
4.
Using the Allen wrench supplied, secure pole clamp to monitor with the set screw
provided.
5.
Replace black knob screw.
6.
To attach to pole, place pole within clamp bracket and tighten screw using the black
finger knob.
4.1.4.2
Optional Mounting Accessories
For information on optional mounting accessories, request Aspect’s “Monitor Mounting
Solutions” booklet (part number 070-0031).
4.2
Instrument Connections
Detailed connection instructions are provided in the BIS VISTA Operating Manual.
4.2.1
Connecting the BISx
1.
Connect the BISx to the monitor
Holding the cylindrical connector with the flat side up, plug the BISx Monitor
Interface Cable into the BISx port on the front of the monitor.
Once connected, the BISx need not be disconnected again. However, if you wish to
disconnect the BISx cable from the monitor, carefully grasp the connector and pull.
DO NOT pull on the cable.
2.
Connect the PIC to the BISx
Attach the gray connector of the Patient Interface Cable to the BISx.
Note:
Connect with the BIS logo facing up for proper pin alignment. To disconnect the PIC, grasp
the connector housing and pull firmly. DO NOT pull apart by the cable wire.
4.2.2
Power Cord Connections
The BIS VISTA Monitoring System is designed to use only 3-conductor IEC hospital-grade
power cords. Cords must be type SJE, SJT, or SJO. Check for a firm connection.
4.3
Installation and Verification Procedure
1.
Open packages and inspect for all components:
•
Monitor
•
Power cord
•
Pole clamp
•
BISx
•
PIC (Patient interface cable, connects BISx to patient)
2.
Connect power cable to monitor, plug power plug into appropriate wall outlet.