![Aspect BIS VISTA Service Information Manual Download Page 33](http://html1.mh-extra.com/html/aspect/bis-vista/bis-vista_service-information-manual_2984718033.webp)
SECTION 4 PREPARATION FOR USE AND INSTALLATION
_______________________________________________________________________
4-1
SECTION 4
4
PREPARATION FOR USE AND
INSTALLATION
INTRODUCTION
This section provides an overview of installation information for service personnel working
with the Aspect BIS VISTA Monitoring System. Please see the BIS VISTA Monitoring
System Operating Manual for full installation instructions.
•
Environment
•
Instrument connections
•
Installation and verification procedure
4.1
Environment
4.1.1
Shipping and Storage Environment
The monitor and its accessories can be stored or shipped within the following environmental
limits. Note that these limits apply to non-operational storage and shipping situations.
Temperature
-10
°
C to +60
°
C
Humidity
15% to 95% (non-condensing)
Pressure
360 mmHg to 800 mmHg
Protect the monitor from sudden temperature changes that can lead to condensation within
the instrument. To minimize condensation, avoid moving the system between heated
buildings and outside storage. Once moved inside, allow the monitor to stabilize in the
unopened shipping container at the inside ambient temperature before unpacking and
placing into service. Before operation, wipe down all visible condensation and allow the
system to reach equilibrium at room temperature.
4.1.2
Operating Environment
The BIS VISTA Monitoring System is not designed for use in areas containing flammable
gases or vapors.
WARNING!
EXPLOSION HAZARD: DO NOT USE THE BIS VISTA SYSTEM
IN A FLAMMABLE ATMOSPHERE OR WHERE
CONCENTRATIONS OF FLAMMABLE ANESTHETICS MAY
OCCUR.
MONITOR IS NOT DESIGNED FOR USE IN MRI
ENVIRONMENT.
Temperature:
The BIS VISTA monitor is designed to operate safely at a room temperature
of 0
°
C to 40
°
C. Conditions that exceed these limits could affect reliability.