background image

of 20

6

Pressure values produced by the pump are able to be 

clearly read to within ±0.25% accuracy of the span of the 

precision test gauge selected for testing. Certification of 

test gauge accuracy traceable to the N.I.S.T., is available 

upon request. A hand jack set, three wrenches, spare 

O-rings, and special adapter for making connections to 

pressure outlets that do not accommodate cone pipe seat-

ing, are included with each unit.
Test pump accessories include lower and back connection 

offset pipe assemblies, with pipe adapters for 

1

4

 and 

1

2

 

NPT connections.
An all metal, double-latched, top handle carrying case is 

supplied with the complete tester to hold all fittings and 

attachments.
2.2.2 Theory of Operation – 1327D 

The operating principle behind the 1327D Test Pump is 

a simple form of comparison. Pressure produced by the 

pump is equally distributed by the manifold to a test gauge 

and to a device being calibrated or checked. The reading 

of the test gauge serves as the reference to which other 

device readings are compared against.
Selection of the test gauge range is the determining factor 

in establishing the precision to which a comparison check 

is to be made. Test gauge accuracy is expressed as a per-

centage of its span.
Proper selection of the test gauge range must be made 

to minimize the amount of the unit error. If a 5,000 psi 

(35,000kPa) range test gauge were used to test devices 

whose span values were less than a fourth of the test 

gauge range, the unit error of 12.5 psi (87kPa) would be 

greater than 1% of the test pressure. Therefore, it is im-

portant to select a test gauge that has a full scale range 

equal to or only slightly in excess of the pressure value to 

be measured.
2.2.3 Specifications – 1327D 

Accuracy: 

1

4

% of 1% of test gauge span.

Test Gauges: Ashcroft solid front type 45-1082-02L-XAZ, 

4

1

2

˝ dial size (120mm case). Bourdon tube materials for 

1327D and DH are bronze and Monel. Bourdon tube ma-

terials for 1327DO are bronze and Monel.
Pump: Two stage, lever operated generates 10,000 psi 

(70,000kPa) with 28 pounds (12.7kg) of force on lever 

handle.
Pump Body: Aluminum, corrosion inhibited and coated 

with baked blue epoxy finish.
Shuttle Valve: Stainless steel bypass valve that controls 

rate of pressure increase and reduces operator effort 

when working at high pressure.
Displacement Valve: A fine pitched threaded valve road 

permitting vernier adjustments to fluid volume to provide 

precise changes. Limit stops prevent road removal during 

normal operation.
Mounting: Four bench mounting holes located in base for 

positive mounting to any level surface.
Instrument Connections: Two coned pipe assemblies 

provide vertical calibration capability for back and lower 

connected gauges. Standard 

1

4

 inch internal NPT and 

1

2

 

inch internal NPT fitting adapters are supplied.
Operating Fluid: 1327D – Light grade machine oils, au-

tomotive petroleum base SAE 20 oils or other equivalent 

fluids suitable for use with Buna N O-ring materials. 1.5 

pints required (.7 liters).

1327DH – Most hydraulic oils of phosphate ester base, 

brake fluids, skydrol, pydraul etc., suitable for use with 

Butyl or Ethylene-Propylene O-ring materials. 1.5 pints 

required (.7 liters).
1327DO – Distilled water with compatible corrosion in-

hibitor. A .1% solution of sodium dichromate is a suitable 

inhibitor. 1.5 pints required (.7 liters).
2.2.4 Certification: 

Standard – Master gauges are calibrated on traceable 

standards, no certification documentation provided.
Optional (CD-4) – Traceable calibration certificate for 

each master gauge.
3.0 

SETTING UP THE TESTER

3.1 Inspection
3.1.1 

Component Check 1305D 

The deadweight tester, as shipped from the factory, 

contains all of the components necessary for operation 

except pump fluid. Depending upon the model and pres-

sure range selected, the total number of weights, weight 

boxes and piston and cylinder assemblies will vary. Refer 

to Specifications, Section 2.1, Table 2-1 for model clarifi-

cation.
3.1.2 Component Check 1327D 

The portable test pump, as shipped from the factory, con-

tains all of the components necessary for operation except 

pump fluid. Depending upon the model and pressure 

range selected, the total number and their ranges will vary. 

Refer to Specifications , Section 2.2, Table 2-3 for model 

clarification.
3.1.3 Claim for Missing or Damaged Goods 

In the event it is determined that an item is missing or 

damaged, contact Customer Service immediately at the 

address indicated in Section 1.0. Damaged merchandise 

and the packaging materials must be held for inspection 

by authorized Ashcroft Inc. personnel or the responsible 

carrier. Disposition of damaged or missing goods will be 

determined after inspection has been  performed on loca-

tion, or at the discretion of the manufacturer, at the factory. 

Inspection may be waived if circumstances, as judged by 

the manufacturer, preclude its necessity.
3.2 Installation Type 1305D
3.2.1 Set-up Procedure (See Figure 2-1)
 

To remove the tester from the case, release the spring 

clip, depressing the clip by hand, grasp the pump reservoir 

and lift up the back end of the pump assembly.
Having removed the pump assembly from its carrying 

case, remove the two plastic shipping plugs.
Select a lower connection (64) or back connection (66) off-

set pipe and connect to inner pump body outlet as shown 

in Fig. 2-1.
Remove the reservoir filler plug (35) and fill reservoir with 

operating fluid.
3.2.2 Bench Space Required (See Figure 1-1) 

Allow a space: 

10 inches (25.4 cm) width 

27 inches (68.6 cm) depth 

22 inches (55.9 cm) height – allows for maximum load of 

deadweights
3.2.3 Storage Space Required 

One Tester Box: 

10

1

2

 inch (2.6 cm); 22

1

2

 inch (57 cm) depth; 10 inch (25 

cm) height 

Summary of Contents for 1305D

Page 1: ... Manual for the ASHCROFT Type 1305D Deadweight Tester andType 1327D Portable Pump Installation and Maintenance Manual for the ASHCROFT Type 1305D Deadweight Tester andType 1327D Portable Pump I M002 10095_RevF_ 250 1526 _09 20 22 ...

Page 2: ...of 20 2 ...

Page 3: ...0018 New York NY 10017 1 Pressure Select a range so that the maximum ap plied pressure will never exceed the upper range limit 2 Vibration Excessive vibration could cause a loosen ing of components and abnormal wear resulting in loss of instrument accuracy or failure to provide valid data 3 Temperature Operation of the instrument in an en vironment where temperatures are in excess of design rating...

Page 4: ...nt mass being acted upon by gravity For this rea son the masses supplied with the tester are stamped with two pressure values the value being contingent on the effective area of the piston and cylinder assembly selected 2 1 3 Specifications 1305D Accuracy Combined tolerance of weights and piston and cylinder assemblies within 0 1 of reading Weight toler ance within 0 05 of mass Piston and cylinder...

Page 5: ...1 2 1 26 175 80 Model Piston Assembly Piston Net Number Pressure Range Piston Area Value Number of Weights by Value Weight L 5 L 1 L 2 L 4 L 5 bar Type Low High Low High Low High H 2 5 H 5 H 10 H 20 H 25 lbs kg 1305DB 20 1305DBH 20 1 20 5 2 5 1 1 2 1 2 63 29 1305DB 100 1305DBH 100 1 20 5 100 5 2 5 1 1 2 1 2 65 30 1305DB 150 0613 in 2 0123 in 2 1305DBH 150 1 30 5 150 3954 0791 5 2 5 1 1 2 1 4 75 34...

Page 6: ...andard 1 4 inch internal NPT and 1 2 inch internal NPT fitting adapters are supplied Operating Fluid 1327D Light grade machine oils au tomotive petroleum base SAE 20 oils or other equivalent fluids suitable for use with Buna N O ring materials 1 5 pints required 7 liters 1327DH Most hydraulic oils of phosphate ester base brake fluids skydrol pydraul etc suitable for use with Butyl or Ethylene Prop...

Page 7: ...DMG 250 1327DMGH 250 X X 0 10 0 40 0 250 40 18 1327DMGO 250 1327DMG 600 1327DMGH 600 X X 0 10 0 40 0 250 0 600 42 19 1327DMGO 600 1327DBG 10 1327DBGH 10 X X 0 10 36 16 1327DBGO 10 1327DBG 40 1327DBGH 40 X X 0 10 0 40 38 17 1327DBGO 40 bar 1327DBG 250 1327DBGH 250 X X 0 10 0 40 0 250 40 18 1327DBGO 250 1327DBG 600 1327DBGH 600 X X 0 10 0 40 0 250 0 600 42 19 1327DBGO 600 1327DAG 1000 1327DAGH 1000 ...

Page 8: ...ndle mo tion Repeat this action until no air bubbles are observed in the fluid flow Note the bleed screw must be closed when the handle is being raised Push the shuttle valve in towards the pump body and continue pumping to the desired test pressure Open the release valve 37 to vent the pressure To check operation pull the shuttle valve outward close the release valve and operate the pump handle s...

Page 9: ...of 20 9 Figure 2 1 Ashcroft Portable Deadweight Tester 1305D ...

Page 10: ...he gauge to the desired position and re tightening nut adapter Repeat the above procedure to connect the gauge being tested to the offset pipe assembly 64 or 66 3 4 3 Bench Space Required Allow a space 10 inch 2 5 cm width 23 inch 5 8 cm depth 22 inch 5 6 cm height allows for testing gauges up to 8 5 inch 250 mm size 3 4 4 Storage Space Required 101 2 inch 27 cm width 221 2 in 57 cm depth 10 inch ...

Page 11: ...ght of oil for various pressures A lighter oil may be used where low pressures are being checked a heavier oil where higher pressures are being tested The reservoir can be refilled while the test pump is operat ing under pressure 3 5 7 Caution Standard testers designed for oil service may not be used with water for oxygen service 3 5 8 Hydraulic service testers should not be filled with water or a...

Page 12: ...uracy of the finished weights better than 1 20th of 1 To maintain deadweight tester accuracy handle the weights with care and keep the piston and cyl inder in clean condition The tester should be flushed with a solvent occasionally preferably every six months so that operating fluid is al ways clean This will prolong component life and provide protection against possible sticky action between the ...

Page 13: ...f the tester body assembly casting When replacing O ring packings coat them with suitable lubricant to pre vent sticking and tearing during assembly and tightening of connection 5 14To Reassemble Body assembly on drip pans Replace body assembly 2 on drip pan 3 using bolts 25 5 15 Replacing Displacement Valve a Place O ring packing between back up washers b Attach back up washers and O ring packing...

Page 14: ...of 20 14 Figure 2 2 Ashcroft Portable Test Pump Type 1327D ...

Page 15: ...ylinder 3 Damaged piston plate damaged cylinder threads or damage to any component part that results in inferior performance or malfunctioning of the unit The piston cylinder assembly is available as a unit only This assures the user of maximum accuracy in his Ashcroft Deadweight Tester and maintains certified trace ability to the N I S T 6 0 WARRANTY SHIPPING INSTRUCTIONS HOWTO ORDER 6 1 Warranty...

Page 16: ...ate with deadweights drops rapidly Unable to hold constant pressure Worn piston and cylinder assembly Use a heavy oil temporarily Replace piston cylinder Stacked dead weights wobble when spinning piston plate Damaged dead weights Check dead weights for visible damage bends dents nicks etc and alignment Piston plate as sembly will not spin Too heavy an oil grade being used Flush fill unit with prop...

Page 17: ...of 20 17 4A ...

Page 18: ...d Valve 27 70 52 1 LEE236 Nameplate 4 70 58 1 LK227 Pipe Extension Assembly 180 00 Qty Unit Item Per List No Asmy Part No DescriptionPrice 59 2 LD227A Bushing2 9 50 60 2 LN141 Adapter 25 40 61 4 LD186 Collar 9 50 62 4 LD117 Nut Adapter 25 20 63 1 LH227A Extension Pipe 67 00 64 1 LJ227 Lower Connection Pipe Assembly 65 1 LF227A Lower Connection Pipe 104 00 66 1 LG227A Back Connection Pipe 104 00 67...

Page 19: ... psig 100 500 PSI WEIGHT SETS Weight Set Number of Weights with by High Pressure Value Part Carrying 25 50 100 200 500 Number Case psi psi psi psi psi LBH199C 1000 psi 1 3 2 3 LBH199B 2000 psi 1 3 2 3 2 LBH199A 3000 psi 1 3 2 3 4 LBH199 5000 psi 1 3 2 3 8 LBH199D 5000 to 10 10000 psi For complete 10 000 psi weight set order both part no LBH199 and LBH199D AVAILABLE WEIGHTS WEIGHT SETS ...

Page 20: ...change without notice All sales subject to standard terms and conditions Ashcroft Inc I M002 10095_RevF_ 250 1526 _09 20 22 Ashcroft Inc 250 East Main Street Stratford CT 06614 5145 U S A Tel 203 378 8281 email info ashcroft com www ashcroft com ...

Reviews: