ASG DTT-L 10 User Manual Download Page 8

ASG, Division of Jergens, Inc.  

15700

 S. Waterloo Road | Cleveland, OH 44110-3898 | Phone: (888) 486-6163 | Fax: (216) 481-4519 | Email: [email protected] | www.asg-jergens.com

Page 8 of 15

Battery Indicator

Basic Function

Display of Clockwise/Counter-Clockwise

Battery

Basic Function

TRACK

0.003

kgf.cm

CW

005

TRACK

0.003

kgf.cm

CCW

005

F-PEAK

0.003

kgf.cm

CCW

005

PEAK

0.003

kgf.cm

CCW

005

Battery Level > 5.00V

4.99V > Battery Level > 4.8V

4.79V > Battery Level > 4.7V

4.69V > Battery Level > 4.65V

Battery Level < 4.65V

If the battery level is less than 4.6 V, the “

Battery Empty

” message will be displayed and the tester 

will power down automatically.  

NOTE:

 Only use the 9V adapter/charger supplied.

Clockwise torque is displayed on the DTT-L and recognized by the symbol “CW.”  Counter-clockwise 
torque is displayed on the DTT-L and recognized by the symbol “CCW.” 

Clockwise symbol

Charging symbol

Load indicator bar 
for clockwise

Load indicator 
bar for counter-
clockwise

Counter-clockwise symbol

A load indicator bar alerts the operator as to how much torque load is being applied to the 
transducer as related to its full scale rating.

DTT-L Series Digital Torque Testers 

User Manual

Summary of Contents for DTT-L 10

Page 1: ...oo Road Cleveland OH 44110 3898 Phone 888 486 6163 Fax 216 481 4519 Email asginfo asg jergens com Web www asg jergens com Page 1 of 15 2017 Jergens Inc All Rights Reserved Revision Date 02 19 18 DTT L...

Page 2: ...ver Adapter 6 Usage 7 Procedure for Manual Torque Screwdrivers 7 Procedure for Power Screwdrivers 7 Battery Indicator 8 Basic Function 8 Display of Clockwise Counter Clockwise 8 Zeroing the Tester 9 C...

Page 3: ...d or implied and all implied warranties of merchantability and fitness for a particular purpose are disclaimed and excluded from this transaction and shall not apply to the goods sold hereunder The AS...

Page 4: ...ble 9 Volt 670 mA 90 264 VDC ASG 66619 ADC 16 Bit CPU 8 Bit Key Button 8 Buttons Unit of Measure 6 7 User Selectable Torque Units Depending on Model Mode of Measure 3 Modes Track Peak and First Peak D...

Page 5: ...eters are supplied with the following standard items DTT L torque meter Battery charger Power adapter NOTE Rundown adapters sold separately International charger also sold separately DTT L Series Digi...

Page 6: ...4 5 26 0 0 51 2 94 66618 1 4 Hex 25 0 80 0 2 8 9 03 66639 5 mm Hex 25 0 80 0 2 8 9 03 66634 3 8 Hex 50 0 200 0 5 65 22 6 66642 3 8 Hex 50 0 500 0 5 65 56 5 Below is a list of all rundown adapters ASG...

Page 7: ...to be sure the readings are consistent The proper reading may be an average of several readings Procedure for Power Screwdrivers Put the appropriate rundown adapter into the unit s female 3 8 square...

Page 8: ...tery Level 4 65V If the battery level is less than 4 6 V the Battery Empty message will be displayed and the tester will power down automatically NOTE Only use the 9V adapter charger supplied Clockwis...

Page 9: ...m lbf in and lbf ft To change the display units press the UNIT key Each time the UNIT key is pressed it will select the next available unit until the tester returns to its original setting The DTT L...

Page 10: ...ROW keys to move the cursor to SETTING Press ENTER To move between the submenus listed in the SETTING menu press the UP or DOWN ARROW keys to move the cursor The sub menus in the setting menu are auto...

Page 11: ...menu page Use the UP and DOWNARROW keys to change the value Use the RIGHTARROW key to change the unit Press ZERO key to reset value to zero Press ENTER key to set new value and return to setting menu...

Page 12: ...To access the Pass Fail menu press the MENU ESC key the display will show the main menu page Use the UP and DOWN ARROW keys to move the cursor point to SETTING Press the ENTER key and use the UP and D...

Page 13: ...ess the MENU ESC key the display will show the main menu page Use the UP and DOWN ARROW keys to move the cursor point to MEMORY Press the ENTER key Press the UP and DOWN ARROW keys to move the cursor...

Page 14: ...key the meter will return to the memory menu page If you select YES and press the ENTER key the meter will delete all saved records and return to the memory menu page DTT L Series Digital Torque Teste...

Page 15: ...o access the Transducer State menu press the MENU ESC key the display will show main menu page Use the UP and DOWN ARROW keys to move the cursor point to TRANSDUCER STATE Press the ENTER key Press the...

Reviews: