Page 12 of 15
Main Menu (Continued)
The pass-fail feature is used to set a defined acceptable maximum and minimum
torque range for measuring. It is activated by setting the lower level and upper level
torque limit. As long as the torque value is within this range, the green LED display
button will light up OK. Any reading values outside this range (higher or lower), the
orange LED display buttons will light up. The orange up arrow will light up if the
torque range is higher than the set range and the orange down arrow will light up if
the torque range is lower than the set range. If this feature is activated, the PF symbol
will display on the main display.
• To access the Pass-Fail menu, press the
MENU/ESC
key, the display will show the
main menu page. Use the
UP
and
DOWN
ARROW
keys to move the cursor point
to
SETTING
. Press the
ENTER
key and use the
UP
and
DOWN ARROW
keys to
move the cursor point to
PASS-FAIL
. Press
ENTER
. The display will show the
Pass-Fail menu page.
• Press the
ESC
key to return the main menu page
• Use the
UP
and
DOWN ARROW
keys to change the value. Use the
RIGHT
ARROW
key to change the unit. Use the
LEFT
ARROW
key to toggle between
the upper and lower limits.
Press the
ZERO
key to reset value to zero. Press the
ENTER
key to set new value and return to setting menu. Press
ESC
to return to
setting menu.
NOTE:
Lower level must be less than the upper level
Pass-Fail
DTT-L Series Digital Torque Testers
User Manual
Example:
Lower Level
Upper Level
= 0 N.m
= 20 N.m
Upper Torque
Level
Lower Torque
Level
Upper Torque
Level
Lower Torque
Level
LOAD
LOAD
LOAD
TIME
TIME
TIME
Example:
Lower Level
Upper Level
= 20 N.m
= 0 N.m
Example:
Lower Level
Upper Level
= 10 N.m
= 20 N.m