ASG, Division of Jergens, Inc.
15700
S. Waterloo Road | Cleveland, OH 44110-3898 | Phone: (888) 486-6163 | Fax: (216) 481-4519 | Email: [email protected] | www.asg-jergens.com
Page 15 of 15
Main Menu (Continued)
About
This shows the information regarding your tester (model, capacity, firmware revision,
serial number).
• To access the About menu, press the
MENU/ESC
key, the display will show main
menu page. Use the
UP
and
DOWN
ARROW
keys to move the cursor point to
ABOUT
. Press the
ENTER
key. The display will show the
ABOUT
menu page.
• Press the
ESC
key to return to the main menu page
Transducer State
This is used to check the status of the transducer. If you suspect that your transducer
has sustained an overload, check the status of the transducer immediately. To
access the Transducer State menu, press the
MENU/ESC
key, the display will show
main menu page. Use the
UP
and
DOWN
ARROW
keys to move the cursor point to
TRANSDUCER STATE
. Press the
ENTER
key. Press the
UP
and
DOWN
ARROW
keys
to move the cursor point to
VIEW
and press the
ENTER
key.
• Place the meter horizontally on a flat level surface and go to the main menu page
• Use the
UP
and
DOWN
ARROW
keys to move the cursor point to
TRANSDUCER
STATE
and press the
ENTER
key
• The display will show the
TRANSDUCER STATE
menu page. Press the
ESC
key to
return to the main menu page
• If the % offset is greater than 10%, please contact ASG to arrange for evaluation
• Transducer state offset number is for reference only and does not determine
whether the transducer is yielded
• These values are given only as an indicator – the need for repair may vary
according to the individual characteristics of the transducer
DTT-L Series Digital Torque Testers