Main Menu (Continued)
The Pass-Fail feature is used to set a defined acceptable maximum and minimum
torque range for measuring. It is activated by setting the lower level and upper
level torque limit. As long as the torque value is within this range, the display will
show a
PASS
message. Any reading values outside this range (higher or lower), the
display will show a
FAIL
message. If this feature is activated, the PF symbol will
display on the main display.
• To access the Pass-Fail menu, press and hold the
MENU
key to access the main
menu. Use the
UP
or
DOWN
arrow keys to move the cursor to
SETTING
and
press
ENTER
. Use the
UP
or
DOWN
arrow keys to move the cursor to
PASS-FAIL
and press
ENTER.
• Press and hold the
DOWN
arrow key to toggle between the upper and lower
limits. Press and hold the
UP
arrow key to change units. Press the
UP
and
DOWN
keys to change the numerics. Press and hold the
SAVE
key to reset
value to 0 to select a new value. Press
ESC
key to return to the setting menu.
• To reset, press and hold the
RESET
button
NOTE:
Pass-Fail feature will be automatically disabled if you set
LOWER LEVEL and
UPPER LEVEL
= 0.000
NOTE:
Lower level must be less than the upper level
Pass-Fail
DTT Series Digital Torque Testers
User Manual
Example:
Lower Level
Upper Level
= 0 N.m
= 20 N.m
Upper Torque
Level
Lower Torque
Level
Upper Torque
Level
Lower Torque
Level
LOAD
LOAD
LOAD
TIME
TIME
TIME
Example:
Lower Level
Upper Level
= 20 N.m
= 0 N.m
Example:
Lower Level
Upper Level
= 10 N.m
= 20 N.m