User manual for ASIN Aqua
17
2.
From the start screen of your
ASIN Aqua select SETTINGS
and then CALIBRATION .
3.
Select the blue pH button on
the right of the screen .
4.
Compare the reading given on
your ASIN Aqua with the reading
taken manually . If different adjust
the reading on the ASIN Aqua by
pressing the + and – buttons so that
they match and press OK to save .
R E CO M M E N DAT I O N:
for precision calibrating use the test buffer pH 7,00 .
If the manually test value is
outside the pH range 6,2 – 7,8
the following notification is given .
If the difference between manually
test and current displayed pH
valve is greater than 0,3 the
following notification is given .
pH buffer 7,00 #12065