EN
32
In Operation Measurement
and Calibration
CLF Probe Calibration
Calibration rules:
Calibration is performed only for refining the value measured by the
colorimeter or Pool Tester and the measured value displayed on the ASIN
AQUA Home (if the difference is up to 0.2, calibration is not necessary).
The probe can be calibrated when the chlorine content in pool water
corresponds to the required value or it is higher (measured with a
colorimeter or Pool Tester).
Use a colorimeter or tester to measure the chlorine value in pool water. The
chlorine concentration must be 0.3 to 1.2 mg/l, otherwise the calibration
cannot be performed.
Calibration procedure :
Compare the value displayed on the ASIN AQUA Home with the one
measured with colorimeter or Pool Tester.
1.
If the difference in values is up to 0.2 mg/l, calibration is not necessary
2.
If the difference in values is
higher
than 0.2 mg/l, you can calibrate the
CLF probe. Go to the CLF calibration menu, set the value measured by
the colorimeter or tester and save it.
3.
If the difference in values is
lower
than 0.2 mg/l, the CLF probe cannot
be calibrated.
Proceed as follows:
•
Go to the menu required value, increase the value by 0.1 mg/l (by 0.2
mg/l max) compared to the current disinfection setting (regardless of the
required value according to the table).
•
After the water in the pool is mixed thoroughly and the value shown on
the ASIN AQUA Home display is settled, repeat the measurement.
•
Repeat the process until the chlorine concentration in pool water
matches the required value
, then set the correct required value
according to the table. You can then calibrate the CLF probe.
After 24 hours
, it is advisable to check the chlorine value with a colorimeter
or Pool Tester. Repeat this procedure until the chlorine concentration in pool
water matches the required value.
Notification:
the values measured in the water may not match with the value
shown on the display during the first few days before the probe stabilizes.