3
EN
1. NOTES • GUIDELINES • GUARANTEE
1.1. GENERAL SAFETY NOTES
• Observe applicable statutes and regulations, and the notes in these operating instructions, when handling
hazardous materials
• Only use the recirculation filter attachment when it is in proper condition
• Please follow the instructions of the safety experts, technical supervisory services (BG, GAA, pharmacy equip-
ment) of the construction supervision, etc.
• Observe accident prevention regulations and workplace ordinance
• Check the recirculation filter attachment regularly for proper function (exhaust air volume flow and filter satura-
tion by checking the indicators)
1.2. GUARANTEE
The guarantee for this product is agreed between you (the customer) and your dealer (the seller). As the
manufacturer, asecos guarantees the products listed in the operating instructions for a period of 24 months
from the date of delivery. All model safety equipment are subject to a compulsory annual inspection by spe-
cialised staff authorised by the manufacturer. Otherwise the customer’s guarantee claim against the manufac-
turer expires.
1.3. DEVELOPMENT AND MANUFACTURER
asecos GmbH, Weiherfeldsiedlung 16-18, 63584 Gründau, Germany
The type plate with the serial number is located on the outside of the unit’s housing
1.4. PURPOSE OF USE
The unit is used for technical ventilation from Ex zone 2 or no Ex zone of safety cabinets used for storing flam-
mable liquids. The recirculation filter attachment is particularly low-noise and therefore ideal for use directly in
the working environment.
Suitable for cabinet models:
UFA Model
Suitable cabinet groups
UFA.025.030
S-LINE / Q-LINE / V-LINE / UB-LINE / CS.195.105.F / E.195.105.F
UFA.025.030.XL
XL-LINE / S-CLASSIC-30
Set-up requirements
OK
OK
OK
OK
CAUTION:
The unit may only be installed in a non-hazardous area!
The integrated activated carbon filter removes 99.999% of the hydrocarbons from the exhaust air so that
it can be discharged directly into the room air. The pollutant-containing exhaust air stream does not come
into contact with electrical current-carrying components. The unit is therefore suitable for exhausting
cabinets that are defined in EN 14470 Part 1 or DIN 12925 Part 1. An exhaust duct leading to the outside
at a non-hazardous location can therefore be omitted. A CE certificate of conformity (also according to EC
Directive 2014/34/EU - ATEX Directive) is available.