5. MALFUNCTIONS • ACTIONS
a
b
c
d
MALFUNCTION
ACTION
the mains plug is inserted
a
does not illuminate
Check if there is power on the socket in use
the On/Off switch had been
activated
b
does not illuminate
LED defective (replace)
or
defect in the electronics
The light switch had been
activated,
The integrated lighting doesn’t
go on
Replace the light bulb.
c
illuminates during operation
of the workplace
Please check the pressure loss on customer’s exhaust air duct (see these
instructions hazardous material workplace! under ventilation adjustment/
monitoring of the hazardous material workplace) or
defect of the integrated air inlet fan or
heavy soiling of the discharge rails.
3 short beeps sound every 60
seconds
c
shortly lluminates every 20
seconds
Power failure
CAUTION
If your hazardous material workplace has the above or other defects, then please contact your dealer.
6. CONNECTION TO A MEDIA SUPPLY
6.1. CONNECTION IN THE MEDIA DUCT
Below worktop
4×20
TX20
1
20 cm
2
3
G 1/2
4
6.2. CONNECTION IN THE MEDIA SOCKET
On backpanel
1
G 1/2
2
3