6
Verifying Speed
To verify centrifuge is running at the correct speed, place a white dot on the cap
of a carrier, insert the carrier in the centrifuge and turn it on.
Point an optical tachometer at the spinning cap and check the reading against
the rotating speed of 3450±100 RPM.
Cleaning Centrifuge, Rotor & Carrier
The user is responsible for maintaining the cleanliness of the centrifuge.
Unplug unit before cleaning or moving.
Any spills must be immediately cleaned up.
If any liquids or debris are spilled on and/or in the rotor the entire centrifuge must
be replaced. Call Ascend Clinical Client Services for a replacement centrifuge:
800.800.5655, Option 1.
Cleaning Centrifuge
Clean the centrifuge weekly using a mild detergent. Spills should be wiped up
immediately.
Clean both the exterior and the interior.
Interior cleaning includes the interior bucket, specimen holder and rubber
cushion.
Disinfect using a 10% bleach solution (one part bleach to nine parts water).
Exterior of the centrifuge should be disinfected weekly.
Interior should be disinfected monthly.
Prepare a fresh 10% bleach solution each time. Once mixed with water, bleach
loses effectiveness after 24 hours. The solution should not be stored for future
use.
Cleaning Rotor
Since all samples are placed into the polycarbonate carrier there is no need to
clean the rotor.
If cleaning the rotor is required please call Ascend Clinical Client Services at
800.800.5655, Option 1 and we will provide a new centrifuge.
Cleaning Carrier
Polycarbonate carrier must be cleaned as needed.
The carrier should be disinfected by autoclave or by soaking in a bleach
containing solution as needed.