background image

 

19)

 

Injection pressure must not exceed the maximum injection pressure when injecting contrast 
media. Exceeding the injection pressure beyond the maximum injection pressure may cause 
damage to the microcatheter.  

20)

 

When infusing contrast media, the device must be operated with confirming that the contrast 
media is being infused from the tip of the device. If the contrast media is not being infused, 
infusion must be stopped and the device must be replaced with a new one.(If the device lumen is 
occluded, the device may be dilated, damaged, or ruptured even at not more than the maximum 
injection pressure, resulting in a life-threatening adverse event due to spurting contrast media.) 

21)

 

Do not use guide wires larger than the recommended size. (Resistance may be felt while 
advancing or withdrawing a guide wire larger than the recommended size, which may cause the 
microcatheter to become damaged or break, or the blood vessel to become damaged. In the 
worst case, life-threatening adverse events may occur.) 

22)

 

If the microcatheter is inserted into vessels and the guide wire is to be replaced, insert the guide 
wire carefully. If there is any resistance during operation, the operation must be discontinued 
immediately, and the microcatheter and the guide wire are withdrawn together. (The 
microcatheter may be damaged and the tip may be cut.) 

23)

 

Repeated insertion and withdrawal of this microcatheter may lead to deterioration of the 
hydrophilic coating. (Continuous use of the device with deteriorated hydrophilic coating may 
cause vascular damage. This may also increase the risk of trapped tip, resulting in a life-
threatening adverse event due to a damage and/or separation of tip.) 

24)

 

Comply with instructions, precautions, and warnings described in the Instructions for Use 
supplied with medical devices (Namely, Sheath introducer kit, Angiographic catheter, Guiding 
catheter, Guide wire) and contrast media used together with the microcatheter.  

 

Precautions

 

1)

 

Do not use if the package is opened or damaged. Always open the package just prior to use. 

2)

 

Prior to use, check all devices, including this microcatheter, and confirm that they function 
normally. Check also if the microcatheter is not damaged during transportation. Do not use if 
the package and/or the microcatheter is suspected to be damaged.  

3)

 

Use immediately after opening the bag. After use, discard it respecting the disposal policies and 
infection controls.  

4)

 

When inserting the guide wire into the microcatheter which is already placed in the blood vessel, 
carefully operate the guide wire not to damage the microcatheter at the bend sections.  

5)

 

Agents containing organic solvent such as alcohol must not be used either alone or concurrently. 
These agents must not be used for immersing or wiping the device. (The catheter may be 
damaged or lose its lubricity.) 

6)

 

Confirm that this microcatheter does not have a kink, knot, torsion, or occlusion before injecting 
contrast media.   

7)

 

Do not use this microcatheter for the purposes other than described in the Indications for Use 
written in this document.  

8)

 

Select the appropriate size of guiding catheter and guide wire to use in combination with this 
microcatheter. (See the label of product package)  

9)

 

When using a guiding catheter fitted with a stopcock, do not manipulate the stopcock after 
inserting this microcatheter into the guiding catheter. (If the stopcock is manipulated during the 
insertion of this microcatheter, this microcatheter may be broken.)  

10)

 

Operate the microcatheter carefully to avoid damage, kinking, or bending, especially when 
inserting this microcatheter into the guiding catheter.  

11)

 

Check the patient’s condition before the procedure. Provide appropriate anticoagulant therapy if 
it is necessary.  

Summary of Contents for Corsair Armet

Page 1: ...ASAHI Corsair Armet Microcatheter SYMBOLS 1 English EN 2 ASAHI Corsair Armet Microcatheter Instructions for Use AMK DT376 Ver 1 11 15TS053...

Page 2: ...away from sunlight Do not reuse Do not resterilize Do not use if package is damaged Caution consult accompanying documents Inner diameter Maximum injection pressure Recommended guide wire diameter Au...

Page 3: ...de support to facilitate the placement of guide wires in the peripheral vasculature and can be used to exchange one guide wire for another The ASAHI Corsair Armet is also intended to assist in the del...

Page 4: ...ways be operated under high resolution fluoroscopic guidance Particular attention should be paid when inserting or withdrawing the microcatheter into stenotic areas and narrower vessels than the produ...

Page 5: ...ctions for Use supplied with medical devices Namely Sheath introducer kit Angiographic catheter Guiding catheter Guide wire and contrast media used together with the microcatheter Precautions 1 Do not...

Page 6: ...this device may result in patient injury 15 Since this catheter is not designed for use in combination with a power injector do not use it in combination with a power injector 16 Take preventive meas...

Page 7: ...enough guide wire support is not obtained secure tightly both the guide wire and the guiding catheter Then advance the microcatheter slowly along the guide wire and observe the movement of the tip to...

Page 8: ...e hemostatic valve of the hemostatic adaptor Remove this microcatheter while keeping the guide wire stable in the blood vessel When this microcatheter is removed check the position of the guide wire u...

Page 9: ...d on the product label 2016 ASAHI INTECC CO LTD 2016 09 08 1st edition ASAHI INTECC CO LTD 3 100 Akatsuki cho Seto Aichi 489 0071 JAPAN EC REP Emergo Europe Molenstraat 15 2513 BH The Hague The Nether...

Reviews: