ASAHI Corsair Armet Instructions For Use Manual Download Page 4

 

5)

 

Do not modify this microcatheter for any reason. Use of a modified product may cause damage 
to blood vessels and/or accidents. 

6)

 

This microcatheter must be used under the condition that surgical operation can be performed. 
(If an emergency surgical operation is unavailable, life-threatening events may occur.) 

7)

 

This microcatheter must be used under fluoroscopy only by a physician who is fully trained in  
PTA treatment. (Unskillful procedure may cause an error in operation or misjudgment, leading 
to damage to the blood vessel. In the worst case, life-threatening adverse events may occur.) 

8)

 

Do not use this microcatheter for patients who are or could be pregnant. (The fetus may be 
affected by X-rays under fluoroscopy.)  

9)

 

Do not use the microcatheter after the expiration date indicated on the label. Discard any device 
that exceeds the expiration date. 

10)

 

The injection pressure must not exceed the maximum injection pressure. (The maximum 
injection pressure is labeled on the product label.) 

11)

 

Excess rotational load must not be applied if the microcatheter is bent. (The microcatheter may 
be damaged.) 

12)

 

If any resistance or something abnormal is felt when operating this product, do not continue the 
operation while the causes are unclear. If it is suspected that the product is not operating 
correctly, avoid excessive manipulations, and carefully remove the entire catheter system while 
paying full attention to avoid complications. (Continuing the operation while the cause of the 
problem is not identified may cause damage to or separation of the catheter, and damage the 
blood vessel. Life-threatening adverse events may occur.) 

13)

 

The microcatheter must always be operated under high-resolution fluoroscopic guidance. 
Particular attention should be paid when inserting or withdrawing the microcatheter into stenotic 
areas and narrower vessels than the product. (Abrasion may result in damage or separation of 
the microcatheter. This may cause vascular injury and perforation, possibly leading to a life-
threatening adverse event.) 

14)

 

Do not insert the guide wire by force or advance it rapidly when the microcatheter is bent or 
twisted. Such manipulations may cause breakage or damage of the microcatheter, or perforation 
of the blood vessel.  

15)

 

Always advance the guide wire ahead of the microcatheter before attempting any manipulation 
of the microcatheter. (If the guide wire is not advanced ahead of the microcatheter, the blood 
vessel may be damaged or perforated, or the microcatheter may be damaged.)  

16)

 

Always hold the connector with one hand and turn the microcatheter carefully while regularly 
releasing the accumulated torsion of the microcatheter. Never turn the microcatheter 
continuously while holding the connector with both hands or use any other means to apply force. 
When releasing the accumulated torsion, be sure to open the hemostatic valve on the Y-
connector. Do not turn the microcatheter in the same direction, either clockwise or 
counterclockwise, for more than 10 consecutive turns. (Continuing rotation may damage or 
break the microcatheter or damage the blood vessels. In the worst case, life-threatening adverse 
events may occur.) 

17)

 

This microcatheter is coated with hydrophilic coating. Therefore, the microcatheter is highly 
lubricious. Always confirm the position of the distal end of this microcatheter and manipulate 
this microcatheter carefully.  

18)

 

Do not infuse contrast media when the microcatheter is bent or occluded. It may cause damage 
to the microcatheter such as expansion or breakage.  

Summary of Contents for Corsair Armet

Page 1: ...ASAHI Corsair Armet Microcatheter SYMBOLS 1 English EN 2 ASAHI Corsair Armet Microcatheter Instructions for Use AMK DT376 Ver 1 11 15TS053...

Page 2: ...away from sunlight Do not reuse Do not resterilize Do not use if package is damaged Caution consult accompanying documents Inner diameter Maximum injection pressure Recommended guide wire diameter Au...

Page 3: ...de support to facilitate the placement of guide wires in the peripheral vasculature and can be used to exchange one guide wire for another The ASAHI Corsair Armet is also intended to assist in the del...

Page 4: ...ways be operated under high resolution fluoroscopic guidance Particular attention should be paid when inserting or withdrawing the microcatheter into stenotic areas and narrower vessels than the produ...

Page 5: ...ctions for Use supplied with medical devices Namely Sheath introducer kit Angiographic catheter Guiding catheter Guide wire and contrast media used together with the microcatheter Precautions 1 Do not...

Page 6: ...this device may result in patient injury 15 Since this catheter is not designed for use in combination with a power injector do not use it in combination with a power injector 16 Take preventive meas...

Page 7: ...enough guide wire support is not obtained secure tightly both the guide wire and the guiding catheter Then advance the microcatheter slowly along the guide wire and observe the movement of the tip to...

Page 8: ...e hemostatic valve of the hemostatic adaptor Remove this microcatheter while keeping the guide wire stable in the blood vessel When this microcatheter is removed check the position of the guide wire u...

Page 9: ...d on the product label 2016 ASAHI INTECC CO LTD 2016 09 08 1st edition ASAHI INTECC CO LTD 3 100 Akatsuki cho Seto Aichi 489 0071 JAPAN EC REP Emergo Europe Molenstraat 15 2513 BH The Hague The Nether...

Reviews: