AS ONE NDK-1A-F Instruction Manual Download Page 14

- 13 - 

 

-|Transparent setting guide|-

1. Timer mode

Setting the timer

You can set the time for start, progress and end of control output.

Set the timer mode in timer [

ST-T

]of the SETUP group.

when the operation stops, press the 

MODE

 to set the time in the timer mode.

 

※ Setting range: 00:00 to 99:59 (hour:minute)

 

※ The icon of each timer mode flashes every 0.5 seconds during the setting time. 

 

※ The set temperature (SV) cannot be set during the setting time. If you turn the JOG switch,the 

SV on the LCD indicator flashes for 2.5 seconds.

 

※ In case of the setting time is 00:00, 

ERR

 appears on the LCD indicator during the operation.   

 

※ When the operation sound [

BUZ

] is set to 

ON

,the sound is generated once when reaching the 

setting time.

 

※ After completing the setting time, press the   key for 3 seconds to stop the operation. 

 

※ When timer mode is changed in timer [

ST-T

] of SETUP group, the setting time (

ON

OFF

WAIT

) is initialized.

 

※ When using the timer, the program setting [

PRG

] of SETUP group does not appear. 

Mode

Operation

Description

Timer OFF

[

OFF

]

SV

Temp.

Time

The drive and controls start at the same 

time and keeps this work until the power is 

OFF.

Timer ON

[

T--1

]

 

 

SV

Control section

Time

Setting time [

ON

]

Temp.

It starts the drive and counts the setting 

time [

ON

] at the same time.

The control output is OFF during the setting 

time [

ON

].

The control output is ON after the setting 

time [

ON

].

Timer OFF 

[

T--2

]

 

 

SV

Temp.

Control section

Time

Setting time [

OFF

]

It starts the drive and counts the setting 

time [

OFF

] at the same time. 

The control output is ON during the setting 

time [

OFF

].

The control output is OFF after the setting 

time [

OFF

].

Timer 

ON/OFF 

[

T--3

]

 

 

SV

Temp.

Control section

Time

Setting time 1

[

ON

]

Setting time 2

 

[

OFF

]

It starts the drive and counts the setting 

time 1 [

ON

] at the same time, the control 

output is OFF during the setting time 1 [

ON

].

After the setting time 1 [

ON

], the setting 

time 2 [

OFF

] starts and the control output 

is ON during the setting time 2 [

OFF

].

After the setting time 2 [

OFF

], the control 

output is OFF. 

SV 

[

T--4

]

 

 

 

SV

Temp.

Control section

Time

Setting time

[

WAIT

]

The drive and controls start at the same 

time, if the PV is higher than SV, the control 

output is maintained during the setting 

time [

WAIT

] and then the output goes to 

OFF. 

 

Summary of Contents for NDK-1A-F

Page 1: ...ents Safety considerations 1 Caution during use 2 Installation instructions 3 Product overview 4 Order of the operation 7 SETUP group 9 SETUP group function Alarm temperature correction etc 11 Setting...

Page 2: ...aution of high temperature under certain conditions reducing the performance This indicates the risk of electric shock under certain conditions This indicates the caution for explosion risk under cert...

Page 3: ...Use the power cable after confirming that there is no dust or moisture The ambient temperature is 5 to 35 Please do not operate in beyond of the range The ambient humidity is 35 to 85 RH Please use i...

Page 4: ...f equipment Please avoid the place where there is dust or is generated Please avoid the place where there is a shock and vibration Please take an enough space for a test easily Front Rear Left Right m...

Page 5: ...ature of the top plate when the operation stops 2 LCD indicator It displays icons from the setting set temperature SV present temperature PV and the amount of control output 3 JOG switch Press over 3...

Page 6: ...7A 6 4 30mm 3 Alarm output terminal It is a connector to connect alarm output relay contacts wires 4 External thermocouple input terminal It is a connector to connect thermocouple sensor K CA sold sep...

Page 7: ...6 Transparent setting guide Product overview Dimensions Unit mm NDK 1A I NDK 1A F NDK 2A I NDK 2A F 204 204 240 69 5 90 280 284 360 320 284 69 5 90...

Page 8: ...te of stop operation Before using the product 1 You must readcaution during use see page 2 andinstallation instructions see page 3 and make sure that these instructions are followed correctly 2 Please...

Page 9: ...g program PRG in the SETUP group is set to ON See Setting the program It is activated when the timer ST T of SETUP group is set See Setting the timer It displays integrated operating time See Other fu...

Page 10: ...FF T 1 T 2 T 3 OFF T 4 ON STOP mode Setting range 0 0 to 350 0 It appears when the alarm ALM is set to ON Press the SETUP key once during the operation checking the parameter value of the SETUP group...

Page 11: ...ol is activated Setting range 1 120 sec It appears when the PID control is activated Setting range 0 0 100 0 In case of the PID control it appears when the intergral time I is set to 0000 Setting rang...

Page 12: ...set auto tuning AT to OFF It maintains P I D setting values before auto tuning If the sensor break error OPEN occurs during auto tuning auto tuning ends forcibly It maintains P I D setting values befo...

Page 13: ...table The causes of hunting are the setting value of the hysteresis HYS the response characteristics of the controlled subject and the attachment of the sensor In order to minimize this hunting width...

Page 14: ...ion Timer OFF OFF SV Temp Time The drive and controls start at the same time and keeps this work until the power is OFF Timer ON T 1 SV Control section Time Setting time ON Temp It starts the drive an...

Page 15: ...ode Timer ST T of SETUP group is set to OFF Setting the time of timer Turning the JOG switch It changes the setting value Press the JOG switch once Save the setting value and move the parameter After...

Page 16: ...gram operation LCD indicator displays the present temperature PV and the target value of the step SV The icon for each section flashes every 0 5 seconds If you turn the JOG switch during program opera...

Page 17: ...ting value SV1 TIM1 WAT1 Set steps at least 2 to 8 If the run time TIM is set to 0 the step will not proceed from the current step Choose one program among the program 1 to 9 The setting method of pro...

Page 18: ...nt keys and JOG switch can be used LED bar Operation Red Green Run Normal Error Stop Top plate temperature 60 or less Top plate temperature 60 or more ON Flash OFF There is an LED bar on the front of...

Page 19: ...indication part Check the connection of temperature sensor and SV in the stop mode If the message continuously appears check the product Please contact your purchasing place CAL A problem has occurre...

Page 20: ...or damage on the surface And it may cause the discoloration and deterioration of rubber or plastic parts If you use the neutral detergent please clean with a dry towel Always keep the product in dry s...

Page 21: ...per Check there is machinery generates high frequency noise around the product The factory default of temperature control is PID control P 3 0 I 40 D 10 Set suitable PID values for operational conditi...

Page 22: ...3600 sec Derivative time D 0 to 3600 sec Control period T 1 to 120 sec Environment Ambient temp 5 to 35 storage 10 to 50 Ambient humi 35 to 85 RH storage 35 to 85 RH Safety device Fuse 2 bimetal switc...

Page 23: ...installed under the top plate It does not matter that small dust falls Also white and cloth like insulation may be seen but there is no problem in use Q Is it possible to turn on the power or heater...

Page 24: ...23 Transparent setting guide...

Page 25: ...24 Transparent setting guide...

Page 26: ...25 Transparent setting guide...

Page 27: ...ion 2 1 27 Edobori Nishi ku Osaka 550 8527 Japan Tel 81 6 6447 8905 Fax 81 6 6447 8911 E mail asoneint so as 1 co jp The specification and dimensions can be changed without notice in order to improve...