● The [Manual] function is used
to adjust exercise range at the stop mode.
● Touch the [Angle] display area in screen <Fig 4.8> while exercise is stopped,
[Manual] is displayed like screen <Fig 4.9>.
● Pressing and holding the [Up arrow] increase current angle, if it is reached at
desired angle, touch [Upper limit] display area to set as [Upper limit] angle.
The <Fig 4.9> shows the setting being changed from 110° to 130°.
● Pressing and holding the [Down arrow] decrease current angle, if it is reached at
desired angle, touch [Lower limit] display area to set as [Lower limit] angle.
The <Fig 4.10> shows the setting being changed from 50° to 30°.
4.2.7 Locking the touch function
<Fig 4.11>
●
It is possible to lock the touch function in order not to change a programmed value by
patient.
● Pressing and holding the [STOP] button for 5 seconds in the stop mode, the [Lock]
function is set with ‘beep’ sound and symbol displaying, <Fig 4.11> shows this.
●
It is available to operate only [START] and [STOP] button.
●
The [Lock] function is disabled
with ‘beep’ sound and symbol di
sappear
ing
by pressing
and holding
the [STOP] button for 5
s
econds
.