
CST. CO., LTD.
15
2) Oscillation interval
<Fig 4.4>
●
When the device reaches the Programmed [Upper(flexion) or Lower(extension) limit]
angle, it will oscillate between the programmed [Oscillation] angle.
● If touch twice the [Progressive, Oscillation] display area(6) at the screen <Fig 4.1>,
the
screen is changed to <Fig 4.4>.
● Touch the [Angle] or [Counter] display area, and change the angle or the number of
the [Oscillation] with [Up/Down] arrow(4).
● [Angle] is the range of angle to be done the [Oscillation] function at the [Upper and
Lower limit]. [Counter] is the number of oscillation.
● The range between the [Upper and Lower limit] angle should be 2 times larger than
the programmed [Oscillation] angle.
● When the [Oscillation] function is used, the [Upper and Lower limit] angle is adjusted
automatically if the programmed range between [Upper and Lower limit] angle is not
2 times larger than the programmed [Oscillation] angle.
● With the [Oscillation] angle is already programmed, if the range between [Upper and
Lower limit] angle is programmed to less than 3 times the [Oscillation] angle, then the
[Oscillation] function is automatically disabled.
● Setting example
※
It is not available to use consequently both [Progressive] and [Oscillation].
< Oscillation Condition >
Upper limit angle : 140°
Lower limit angle : -10°
Oscillation angle : 15°
Oscillation Counter : 3
Summary of Contents for ARTUS-903K
Page 22: ...CST CO LTD 21 Sep 2017 printed...