DrillSaw Highspeed 200™
User’s Guide
DFU-0225-2 Rev. 0
Page 22 of 58
Set up the device so that the power switch is easily accessible. In dangerous
situations the device can be disconnected from the power supply using the
power switch.
The power switch can also be used to safely stop the device. In dangerous
situations the device can be disconnected from the power supply by
disconnecting the mains-cable.
No excessive heat generation within the operation mode. The responsibility for
the timely switch off of the motor is the responsibility of the user.
W A R N I N G !
Do not have the device in direct contact with the patient if high-frequency devices
are in use, or if the patient requires defibrillation.
4.3
Replacing the Fuses
The main fuse is replaced as follows:
1. Disconnect the device from the AC mains power supply.
2. Open the fuse tray in the AC inlet [10] by pinching the tabs and pulling
outward.
3. Replace the fuses with 2 x 250V – T1.6AH Line Fuses as noted on the rear
panel.
4. Push the fuse holder back into the AC inlet.
5. Ensure that the fuse holder is fully seated and that the tabs snap back.
Always use fuses with the correct values to avoid allowing
overcurrent to enter the system. Only use Arthrex original fuses.
An incorrect fuse may increase the risk of electrical shock or fire
hazard.
The AR-200C console incorporates a universal AC input power supply.
A voltage selection switch is not required.
4.4
Electromagnetic Compatibility (EMC)
This device has passed testing for EMC compatibility. This device
may cause interference with other devices in the near vicinity if
not set up and used as instructed by Arthrex.
The AR-200 has been designed to accept EMC from other devices within the
limitations as described in section 14.