DrillSaw Highspeed 200™
User’s Guide
DFU-0225-2 Rev. 0
Page 2 of 58
14. The ESD (electrostatic discharge) spring contact on the underside of the foot
control must touch the floor during operation.
15.
Do not
open or attempt to service this system, as this may void your
warranty. There are no user-serviceable parts inside. Removing the cover
may introduce an electric shock hazard by exposing you to dangerously high
voltages or other risks. If the system malfunctions, return it for servicing
immediately.
16. Misuse may damage the AR-200 and may cause risks and hazards for
patients, users and third parties.
17. Misuse, as well as unauthorized assembly and modifications or repairs to the
AR-200 system, or non-compliance with our instructions relieves us of all
liability for claims under warranty or other claims.
18.
Use only
Arthrex approved accessories.
Other accessories may result in
increased emissions or decreased immunity of the system. Contact your
Arthrex representative for a complete list of accessories.
19.
Do not
modify the console or any accessory. Failure to comply may result in
injury to the patient and/or operating room staff.
20.
Do not
use improper or damaged accessories.
21.
Do not
use in the presence of flammable anesthetics or oxidizing gases such
as nitrous oxide, oxygen, or endogenous gases. All oxygen connections must
be leak free for the duration of the surgical procedure.
22. After autoclaving, the accessory devices are VERY HOT. Handle with care to
avoid burns.
23. The console, the motors with cable and the footswitch are not suitable for use
in potentially explosive atmospheres or with potentially explosive mixtures of
anesthetic substances containing oxygen or nitrous oxide.
Zone M is defined
as a ”medical environment“ and constitutes the part of a room in which
potentially explosive atmospheres may form, due to the use of anesthetics or
medical antiseptics and antibacterial soaps; such atmospheres are typically
localized and temporary. Zone M comprises a truncated pyramid below the
operating table which is tilted outwards at a 30° angle.
24. Zone G, also known as an ”enclosed medical gas system“, does not
necessarily include areas enclosed around all sides, in which explosive
mixtures are continuously or temporarily generated, directed or used in small
quantities.
25. Never use the Arthrex 200 in areas close to strong magnetic fields.
26. Do not place the handpiece drive on the patient.
The PRECAUTION! symbol identifies methods and procedures
that must be followed to avoid damaging the device or causing it
to malfunction.
1.
Do not
disconnect the plug of the motors or foot control unit by pulling on the
cable. Remove it by grasping and pulling on the body of the connector.