Synergy
HD3
Cable Kit, Arthrex Pump and Shaver
Operation Instructions
Pump, Shaver, Apollo, & Insufflation Setting Display
Synergy
HD3
™
and Synergy
UHD4
™
currently support the ability to display pump, Apollo, Insufflation and
shaver settings from the current Arthrex pump [AR-6480], shaver [AR-8305], insufflator [AR-3290-
0004), and Apollo [AR-9800] on the surgical monitor.
Display of auxiliary device settings on the surgical monitor allows the surgeon to view these settings
at any time in the procedure even if the pump and shaver consoles are not visible to the surgeon.
Requirements: To utilize the pump and shaver setting display a facility needs to:
•
Have the most recent hardware version of the Arthrex pump, shaver, insufflator, and Apollo
that supports Synergy
HD3
™
and Synergy
UHD4
™
integration.
•
Have the Arthrex pump/shaver integration kit (part # AR-3200-1040).
•
Have Synergy
Version x.6 or newer.
Limitations: While auxiliary device settings are displayed on the surgical monitor, there is no control
of the auxiliary device from Synergy. Auxiliary device settings changes must be made directly on the
auxiliary consoles.
Set Up/Configuration: To configure the Synergy for auxiliary settings display, log in as an
Administrator on the Synergy tablet. Enter System Set-Up by pressing the Maintenance Icon located
in the upper left corner of Synergy tablet (as shown in Figure 1) and then select Advanced Settings.
Figure 1. Maintenance icon on Synergy tablet
From the System tab (as shown in Figure 2) scroll down and locate the Enable Pump, Enable
Shaver, Enable RF, and Enable Insufflator settings and turn these on for each device type
connected to Synergy. Once they are on touch the Save button in the lower right corner of the
screen to save the settings.