Troubleshooting Guide (continued)
Symptom or Message
Cause
Corrective Action
Filter wheel error
Press any Key
The instrument is unable to reliably
position the filter wheel.
Call Artel Technical Support for
assistance.
Inst. Limit Reached
Press any Key
The instrument has performed the
maximum number of readings allowed
for a single pipette calibration.
Press any key; the calibration will be
ended. You may continue with another
calibration.
Invalid Blank Rdg.
Retry? (Yes/No)
Wrong vial inserted or blank out of
specification
Ensure that an unused blank vial is
inserted, and expiration date of kit is not
exceeded.
Invalid CAL A/B/C/D
Press any Key
The wrong calibrator vial, or an invalid
calibrator vial, was inserted during an
instrument calibration.
Press any key; the instrument will ask if
you want to insert another vial or end
the instrument calibration.
Invalid Character
Press any key
alternating with
header entry display
An invalid ASCII character value was
entered when setting up the report
header.
Press any key and re-enter the ASCII
character value.
Invalid volume
Re-enter it
Volume entered is outside the range that
the PCS is capable of calibrating. The
PCS can calibrate pipettes from 0.1 µL
through 5000 µL.
Re-enter the pipette volume, making
sure it is within the PCS’ range.
Invalid volume
Re-enter? (Yes/No)
Volume entered is outside the range that
the PCS is capable of calibrating. The
PCS can calibrate pipettes from 0.1 µL
through 5000 µL.
Re-enter the pipette volume, making
sure it is within the PCS’ range.
Invalid # Re-enter
User is attempting to delete an invalid
data point – that is, a point whose
number is higher than the highest data
point obtained.
Re-enter the data point.
Memory Failure!
Press any Key
The user section of EEPROM is
corrupted.
Press any key except RESET, then
follow on-screen instructions (the
instrument will perform a self-
calibration). Check that the printout
header is correct; if not, re-enter it.
Call Artel Technical Support if you need
assistance with these steps.
Memory Protect Error
Press any Key
Factory calibration memory is
unprotected and could be inadvertently
modified.
Call Artel Technical Support for
assistance.
MIXER FAILURE
Press any Key
Vial hitting vial holder cover, due to vial
not being fully inserted in the vial holder.
Make certain vial is fully inserted in the
vial holder.
An obstruction prevents the vial holder
from moving freely.
Remove gray spill shield, and rear
access cover if needed. Find and
remove obstruction. Refer to Section
8.3, “Removal of Rear Access Cover.”
Component failure
Call Artel Technical Support for
assistance.
Summary of Contents for PCS 2
Page 6: ...15A3356O 2 This Page Intentionally Left Blank ...
Page 10: ...15A3356O 6 This Page Intentionally Left Blank ...
Page 48: ...15A3356O 44 This Page Intentionally Left Blank ...
Page 52: ...15A3356O 48 This Page Intentionally Left Blank ...
Page 54: ...15A3356O This Page Intentionally Left Blank ...
Page 60: ...15A3356O This Page Intentionally Left Blank ...
Page 62: ...15A3356O This Page Intentionally Left Blank ...
Page 64: ...15A3356O This Page Intentionally Left Blank ...