685 Series
ComboPak
User’s Manua
l · Page 9
Settings
All parameters of the
ComboPak
can be accessed and changed over the
computer interface.
Most parameters can be changed using a standard command, as defined in the
Computer Interfacing Manual
or through the
Settings
dialog within
Arroyo
Control
. Below is a list of the most common settings and the corresponding
computer command to change it.
Details on how to communicate with the
ComboPak
using your own program
can be found in the
Computer Interfacing Manual
which is included on the CD
that accompanied this product.
Menu
Description
Factory
Default
Computer
Command
Basic Laser Settings
Mode
As described in the
Laser Control
Modes
section below, the unit
offers five control modes: Io (ACC),
Io HiBW (ACC), Im (AMC), Po
(APC), and Vf (AVC). Change this
setting to select a new mode.
Io (ACC)
LAS:MODE
Io limit
This setting controls the maximum
amount of forward current that can
be delivered to the laser diode. This
limit is implemented in hardware for
immediate response. For more
information about limits, see
Using
the Laser Limits
below.
Maximum
LAS:LIM:LDI
Im limit
This setting controls the maximum
amount of monitor photodiode
current the unit will allow. This limit
is implemented in software. For
more information about limits, see
Using the Laser Limits
below.
Maximum
LAS:LIM:MDI
Po limit
This setting controls the maximum
amount of monitor photodiode
power the unit will allow. This limit
is implemented in software. For
more information about limits, see
Using the Laser Limits
below.
Maximum
LAS:LIM:MDP