AMC-1222 CO/NO
2
Multidrop
17
5.2.3 SPAN ADJUSTMENT FOR (CO) 75 PPM (AMC-1222-3)
Step 8
Apply CO span gas, CO span gas concentration should be in 25ppm to 75ppm CO
range
Step 9
After the signal reading has stabilized (approx 3 min) adjust Span potentiometer to
2.00VDC/75*Span gas concentration. i.e. if using 25 ppm CO
2.00VDC/75*25=0.6667VDC
Step 10
If Signal > .7VDC verify CO Alarm LED Flashes, If Signal > 2.00VDC verify CO Alarm
LED Illuminates steady State
Step 11
Remove Gas, Calibration complete
Figure 5-1 Calibration Procedure for 1222 CO