AquaVENT® FD140i, User Manual, English Issue 01
69
Immunity test
IEC 60601-1-2
test level
Compliance
level
Electromagnec environment -
guidance
Portable and mobile RF communicaons equipment should be used no closer to the AquaVENT
®
FD140i,
including cables, than the recommended separaon distance calculated from the equaon applicable to the
frequency of the transmi@er.
Conducted RF
to IEC 61000-4-6
3 V rms
150 kHz to 80
MHz
10 V
d
= 0,35
P
Radiated RF EM fields
EN 61000-4-3:2006
+A1:2008+IS1:2009+A2:2010
3 V/m
80 MHz to 2.7
GHz
3 V/m
Swept frequency tesng to be
performed on 4 faces of the Aqua-
VENT
®
FD140i.
where P is the maximum output
power rang of the transmi@er in
wa@s (W) according to the
transmi@er manufacturer and d is
the recommended separaon
distance in metres (m).
Field strengths from fixed RF
transmi@ers, as determined by an
electromagnec site survey, a
should be less than the
compliance level in each
frequency range. b
In the vicinity of equipment
marked with the following
symbol, interference may occur.
References and footnotes are explained on the following page.
Note 1 At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2 These guidelines may not apply in all situaons. Electromagnec propagaon is affected by absorpon
and reflecon from structures, objects, and people.
a
Field strengths from fixed transmi@ers, such as base staons for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theorecally
with accuracy. To assess the electromagnec environment due to fixed RF transmi@ers, an electromagnec site
survey should be considered. If the measured field strength in the locaon in which the AquaVENT
®
FD140i is
used exceeds the applicable RF compliance level above, the AquaVENT
®
FD140i should be observed to verify
normal operaon. If abnormal performance is observed, addional measures may be necessary, such as re-
orienng or relocang the system.
b
Over the frequency range of 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
Summary of Contents for 10190001
Page 1: ...User Manual English...
Page 3: ...AquaVENT FD140i User Manual English Issue 01 3 Contents...
Page 6: ...AquaVENT FD140i User Manual English Issue 01 6 1 Introduc on...
Page 13: ...AquaVENT FD140i User Manual English Issue 01 13 2 AquaVENT FD140i Overview...
Page 20: ...AquaVENT FD140i User Manual English Issue 01 20 Device setup 3 Device Setup...
Page 33: ...AquaVENT FD140i User Manual English Issue 01 33 4 Using the AquaVENT FD140i...
Page 49: ...AquaVENT FD140i User Manual English Issue 01 49 5 Alarms and No ca ons...
Page 58: ...AquaVENT FD140i User Manual English Issue 01 58 6 Maintenance and Repair...
Page 60: ...AquaVENT FD140i User Manual English Issue 01 60 7 Cleaning and Decontamina on...
Page 62: ...AquaVENT FD140i User Manual English Issue 01 62 8 Technical Speci ca ons...
Page 71: ...AquaVENT FD140i User Manual English Issue 01 71 9 European Community Declara on of Conformity...
Page 73: ...AquaVENT FD140i User Manual English Issue 01 73 10 Disposal...
Page 75: ...AquaVENT FD140i User Manual English Issue 01 75 11 Appendices...