![Arkray GLUCOCARD SM Operating Manual Download Page 6](http://html1.mh-extra.com/html/arkray/glucocard-sm/glucocard-sm_operating-manual_2972863006.webp)
Chapter 8 Control test
This meter checks itself every time you insert a test strip. You do not need to
run frequent control tests. Do a control test if:
You suspect the meter or test strips are not working properly.
You dropped the meter.
You damaged the meter.
Your test results do not match how you feel.
You want to check performance of the meter and test strips before a blood
glucose test.
NOTE
Use only the GLUCOCARD
TM
SM CONTROL to test your meter and test strips.
CAUTION
Do not
use control solutions beyond their expiration date. The expiration
date is written on the control solution bottles next to the “Use by” symbol.
Leave the meter, test strips and control solution at 8 to 40 °C
(46 to 104 °F) and 20 to 80% RH for at least 30 minutes before running
the control test.
Do not
drink control solution. It is not for human consumption.
? $ "
1. See steps 1 and 2 in paragraph 2.7.
screen.
!
the test strip to a drop of control solution.
Accurate test results will not be obtained if you apply the control solution to the
check window directly.
The meter counts down. After 5 seconds, a beep sounds and the test result,
`
NOTE
Tightly close the control solution bottle.
Do not
touch the test strip until the test result appears.
"
{
the measurement results.
CAUTION
+
solutions, and contact your distributor.
5. Check that your test result is within the acceptable range.
The acceptable range is shown on the label of test strip bottles. If within range,
the meter and the test strip are working properly. If out of range, test the control
solution again.
CAUTION
If you still have test results that are out of the acceptable range, stop using
your meter, test strips and control solutions, and contact your distributor.
6. Point the test strip down. Slide the test strip ejector to eject the test
strip.
The meter turns itself off.
Chapter 9 System care
9.1 Storing your system
After use, tightly close the caps of the test strip bottle and control solution bottle
to maintain their quality.
Store your meter, test strips, control solutions and manuals in your carrying
case and keep this in a dry location. The correct storage temperatures are 0 to
50 °C (32 to 122 °F) for the meter, 1 to 30 °C (34 to 86 °F) for test strips, and 2
to 30 °C (36 to 86 °F) for control solutions. Do not freeze. Avoid heat, humidity
and direct sunlight.
CAUTION
To obtain accurate test results:
Do not
use test strips or control solutions if their bottles are broken or
have been left open.
Do not
use the test strips or control solutions beyond their expiration
dates.
9.2 Cleaning your meter and lancing device
Your meter and lancing device do not need special cleaning. If your meter and
lancing device get dirty, wipe them with a soft piece of cloth moistened with
water. To disinfect these items after cleaning, wipe with a soft cloth moistened
with 70% ethanol, 70% isopropanol or diluted household bleach (0.05% sodium
hypochlorite solution).
9.3 Changing the batteries
When the battery symbol appears permanently on the screen, the
batteries are getting low (approximately 100 tests can be performed).
Before using your meter, change the batteries. Date and time settings
will not be cancelled if you insert new batteries within 30 seconds. Past
results remain in the memory even when the batteries are changed.
NOTE
Your meter uses two 1.5V alkaline LR03/AAA batteries. This type of battery is
available in many stores. Keep spare batteries handy at all times.
1.
Make sure the meter is off.
2.
|
3.
Remove the old batteries.
NOTE
Do not
use sharp objects that may damage the meter.
4.
Insert two AAA alkaline batteries in the correct direction.
5.
Close the battery cover.
NOTE
The meter may enter date setting mode the next time it is
turned on. In this event, set the correct date and time.
} ~
Dispose of old batteries according to your local environmental regulations.