Step 5:
Gently touch the drop of the control solution to the tip
of the test strip. The meter will beep indicating it has
received the control solution.
Step 6:
The screen will start to count down. After 6 seconds,
the control test result will appear on the meter display
screen.
Step 7:
Compare the reading on the screen to the Level 1
range printed on the test strip bottle or its package. If
the reading falls outside the Level 1 range printed on
the test strip bottle or its package, see Control Solution
Trouble Shooting, on page 39.
<
<
PM
Level 1: 53 - 79
Level 2: 135 - 203