Press TOTAL and RESET (15 s). The display will show: ‘P’. If it does not show ‘P’, it means that
the meter is blocked. Then call the Technical Service.
Fill an amount of liquid into a calibrated recipient. Maximum of litres: 99.99. Then press RESET
to accept. Write down the liquid volume in litres (or in the corresponding volume to be
measured) and decimals with the maximum resolution it is possible.
The display
shows ‘L 0.00’ with the right digit blinking. The TOTAL key makes the digit move
forward. The RESET key validates it and goes to the next digit. After the fourth digit, accept it
pressing RESET. The calculated calibration factor will be shown blinking. Press RESET to
accept and end the calibration process or press TOTAL to cancel and not to save the
calibration.
Err C: Calculation error · Err P: without pulses
Err L: value 0.00
When it shows 'Err', it means it has not been calibrated correctly. Try calibrating it again.
6.3. Battery replacing operation
The MAGF1 meter works with a lithium button battery, DL2450 or C2450 model. Its estimated
duration is approx. 4 years (according to its application).
To replace the battery, remove the screws (1), open the cover and remove the screws (6) that
are holding the electronic board.
With the help of a little screwdriver, push the battery to one side.
Replace the old battery with the new one. The correct battery position is the negative pole (-),
placed looking at the bottom (touching the board).
ATTENTION
Do not put the battery upside down. This can damage the electronics. It is important
to follow the battery manufacturer's instructions. Do not throw it to the general
rubbish; throw it to a recycling centre or where the batteries are picked up.