Arkon MAGF1 User Manual Download Page 9

Press TOTAL and RESET (15 s). The display will show: ‘P’. If it does not show ‘P’, it means that 

the meter is blocked. Then call the Technical Service. 
Fill an amount of liquid into a calibrated recipient. Maximum of litres: 99.99. Then press RESET 
to  accept.  Write  down  the  liquid  volume  in  litres  (or  in  the  corresponding  volume  to  be 
measured) and decimals with the maximum resolution it is possible. 
The  display 

shows ‘L 0.00’ with the right digit blinking. The TOTAL key makes the digit move 

forward.  The RESET key validates it and goes to the next digit. After the fourth digit, accept it 
pressing  RESET.  The  calculated  calibration  factor  will  be  shown  blinking.  Press  RESET  to 
accept  and  end  the  calibration  process  or  press  TOTAL  to  cancel  and  not  to  save  the 
calibration. 

Err C: Calculation error · Err P: without pulses 
Err L: value 0.00 
When it shows 'Err', it means it has not been calibrated correctly. Try calibrating it again. 
 
6.3. Battery replacing operation 
The  MAGF1 meter works with a lithium button battery, DL2450 or C2450 model. Its estimated 
duration is approx. 4 years (according to its application).  
To replace the battery, remove the screws (1), open the cover and remove the screws (6) that 
are holding the electronic board. 
With the help of a little screwdriver, push the battery to one side. 
Replace the old battery with the new  one. The correct battery position is the negative pole (-), 
placed looking at the bottom (touching the board). 
 

ATTENTION 
Do not put the battery upside down. This can damage the electronics. It is important 
to  follow  the  battery  manufacturer's  instructions.  Do  not  throw  it  to  the  general 
rubbish; throw it to a recycling centre or where the batteries are picked up. 

Summary of Contents for MAGF1

Page 1: ......

Page 2: ...3 WARNINGS 5 4 SECURITY INSTRUCTIONS 6 5 INSTALLATION 7 6 OPERATION INSTRUCTIONS 8 7 METER ASSEMBLY DISASSEMBLY 11 8 MAINTENANCE 12 9 REPAIR 13 10 PROBLEM GUIDE 14 11 APPENDIX 15 11 1 Country of Origin 15 11 2 EU Declaration of Conformity 15 11 3 Warranty 15 11 4 Contact 15 ...

Page 3: ... 805404002 7 DL2450 or CR2450 BUTTON BATTERY 322209002 8 D 138 94x1 78 mm JOINT 020602007 9 D 6x10 mm GEAR MAGNET 322001008 10 MAGF1 GEAR 326002000 11 D 8x45 mm STAINLESS GEAR SHAFT 326001002 12 DIN 6325 D 3x16 mm PIN 804000021 17 D 38x2 5 mm NBR JOINT 803100056 18 MAGF1 ALUMINIUM BODY 326001000 19 D 74x2 mm NBR JOINT 803100076 20 MAGF1 ALUMINIUM METER COVER 326001001 21 M 5x16 mm DIN 912 BLACK SC...

Page 4: ...a PERMANENT memory The MAGF1 model has two keys and a big LCD Display in order to make easier its display Flow 5 110 l min diesel 5 80 l min lubricant Channels 2 Phase out channel 180º Pulses per channel 20 5 pulses x litre and channel Max pressure loss 0 3 bar diesel Max pressure 55 bar Temperature 10 50 ºC Max humidity 95 H R Viscosity 2 2000 cSt Accuracy 0 5 Noise LEQA 70 dBA Repetitivity 0 2 I...

Page 5: ...ible dangers The operation manual must be considered as a part of the electronic meter and keep it for future inquiries during all its working life We suggest keeping it in a dry and protected place The manual reflects the technical situation at the moment of the meter sale and cannot be considered inadequate for the reason of being updated afterwards according to the new experiences The manufactu...

Page 6: ...the possible accidents it is advisable to read the following warnings and cautions carefully 4 1 ATTENTION A wrong meter installation or use can cause serious effects both physical and material It is advisable to install a filter in order to avoid the solid inlet in the measuring chamber 4 2 ATTENTION When transferring flammable liquids hydrocarbon do not smoke in the work environment A FIRE CAN B...

Page 7: ...asier its reading and key handling unscrew the 4 screws 1 with an Allen screwdriver and put the cover in the desired position Before closing the cover check that the joint 8 is well positioned and screw again the screws 1 in its holes 5 3 It is necessary to use sealing components on the installation pipes and threads which are resistant to the products to be transferred 5 4 ATTENTION Use sealing e...

Page 8: ...unters increase both the partial and the total 6 1 2 Partial Counter Deletion To delete the partial counter press only the RESET key when released the display shows 0 00 and it is ready to count again 6 1 3 Total Counter Reading To see the Total counter press the TOTAL key and the display will show the litre total When released the display shows the partial counter again 6 1 4 Serial Number and Ba...

Page 9: ...t to save the calibration Err C Calculation error Err P without pulses Err L value 0 00 When it shows Err it means it has not been calibrated correctly Try calibrating it again 6 3 Battery replacing operation The MAGF1 meter works with a lithium button battery DL2450 or C2450 model Its estimated duration is approx 4 years according to its application To replace the battery remove the screws 1 open...

Page 10: ...e the battery is changed follow the above steps to reassemble the meter top It is not necessary to make any other operation The electronics remembers the total counter and saves the last calibration factor ...

Page 11: ...r 7 2 Before removing the gears take into account where the magnet gear is assembled If you assemble it the other way round the meter will not work correctly The two points in the drawing must be lined up 7 3 For its assembly please follow the above described steps inversely taking into account the correct meter body screw pin and joint placing REMARK Please be careful the magnets must be in the g...

Page 12: ...eter must be cleaned carefully and when it is going to be mounted you have to make sure that it is correctly done Please follow the instructions from the above Meter Assembly Disassembly no 7 Section If you decide to store the meter for a long time clean it conscientiously It will remain protected and ready for a new starting ...

Page 13: ...livery to be repaired If the meters are not used with oil or diesel they must be rinsed as many times as it is necessary enclosing a note indicating the chemical substances supplied with this unit When ordering spare parts make sure that you give the part number and its correct description This will guarantee the correct supply of the requested spare part ...

Page 14: ...inlets in the system clean and replace the gears Low flow impurities in the measuring chamber clean the measuring chamber The display is not switched on dead battery replace the battery please refer to Section 6 3 The litres are passing but the numeration does not count dead battery REED problem replace the battery please refer to Section 6 3 call the TECHNICAL SERVICE Loss of liquid damaged joint...

Page 15: ...Order Confirmation Please see your Resellers contract or www arkon co uk Support section The Warranty sheet is part of the Packing note of any new goods sent For the claim or return procedure please consult our web site www arkon co uk or call the Arkon Flow Systems s r o sales office The warranty does not cover the inadequate use the use out of the specific regulations the negligence the corrosio...

Reviews: