MAGE1 User Guide
23
V1.5 12-1-2022
9.15. Factory Settings
To enter this section, it is necessary to enter the Factory Password.
Modbus
register
Modbus
address
No. of
bytes
Data
type
No. of
decimal
Min
Value
Max Value
Default
Description
Read/
Write
4000
3999
4
Bool
0
0
1
0
Delete
Volume
R/W
4002
4001
4
Bool
0
0
1
0
Delete Batch
R/W
4004
4003
4
Bool
0
0
1
0
Delete OK
(min)
R/W
4006
4005
4
Bool
0
0
1
0
Delete Error
(min)
R/W
4008
4007
4
Bool
0
0
1
0
Flow
Simulation,
0=OFF,
1=ON
R/W
4010
4009
4
Longint
3
0
36 000 000
3600
Simulated
Flow
R/W
4012
4011
4
Longint
0
0
1 000
-
Diameter
R/W
4014
4013
4
Longint
0
0
999999
-
Unit No.
R/W
4016
4015
4
Longint
3
0
36 000 000
-
Flow Qn
R/W
4018
4017
4
Longint
3
0
36 000 000
-
Calibration
Point 1
R/W
4020
4019
4
Longint
3
0
36 000 000
-
Calibration
Point 2
R/W
4022
4021
4
Longint
3
0
36 000 000
-
Calibration
Point 3
R/W
4024
4023
4
Longint
0
-8388608 8388607
-
Calibration
Data 1
R/W
4026
4025
4
Longint
0
-8388608 8388607
-
Calibration
Data 2
R/W
4028
4027
4
Longint
0
-8388608 8388607
-
Calibration
Data 3
R/W
4030
4029
4
Bool
-
0
1
0
Zero Flow
Set
R/W
4032
4031
4
Bool
-
0
1
0
Zero Flow
Erase
R/W
4034
4033
4
Longint
7
0
1000000
0
Zero Flow
Constant
R/W
4036
4035
4
Longint
-
0
6
4
Excitation
frequency,
0=1/60 Hz,
1=1/30 Hz,
2=1/15 Hz,
3=1/5 Hz,
4=1.5625
Hz,
5=3.125 Hz,
6=6.25 Hz
R/W
4038
4037
4
Bool
-
0
1
1
Excitation,
0=OFF,
1=ON
R/W
4040
4039
4
Bool
-
0
1
0
Reserved
R/W
4042
4041
4
Bool
-
0
1
0
Reset
system
R/W