29
Technical Specification
Component Weights
kg (lbs)
Sara
Plus
- non-scale version (complete - without
battery) 73.8 (162.7)
Sara Plus
- scale version (complete - without
battery) 82.9 (182.8)
Sara Plus
- scale version (complete - with scale and
battery) 88 (194)
Maximum total weight of lift (lift + patient) 278
(614)
Battery pack 4.9 (10.8)
Commode seat and frame (option) 4 (8.8)
Commode pan and holder (option) 1.2 (2.6)
Foot support 5 (11)
All slings - check the safe working load on the sling
label
Electrical
Battery type and part number (Rechargeable - lead
acid): SPL3021
Battery capacity: 6Ah
Battery charger part number: SPL3024**
(Note:** indicates relevant country code)
Fuse 15A (Thermal overcurrent circuit breaker
Fuse – PCBA 20A
Fuse – battery 30A
Protection class lift: IP X4
Protection class hand control: IP X7
Lift nominal voltage: 24V DC
Operating force of controls: < 5 N
Medical Equipment:- type B
protection against
electrical shock in accordance with IEC 601-1)
ArjoHuntleigh patient handling products meet the
requirements of Electromagnetic Compatibility
(EMC) as stated in clause 12.5 of the Medical
Devices Directive 93/42/EEC
Although compliant with EMC
requirements there is a remote chance that
close proximity usage may affect over-
sensitive electrical equipment
The symbol IP n
1
n
2
indicates the degree of ingress
protection against solid particles (n
1
) and liquids
(n
2
).
Conforms to IEC 60601-1:2012(ed.3.1), ANSI/
AAMI ES60601-1:2005 / A2:2010, CAN/CSA-
C22.2 No. 60601-1:08 and ISO 10535:2006.
Duty cycle Max volts Max amps
Mast Lift Actuator (sealed electro-hydraulic unit):
10% (2 min/18 min) 24V 20A
‘V’ Chassis Actuator (electro-mechanical unit):
10% (2 min/18 min) 24V 8A
Maximum sound power level
In accordance with ISO 3746 74 dB
(dB re 1pW ± 3dB)
Environment - Operating
Relative humidity: 30% to 75%
Ambient temperature: 10ºC to 40ºC (50ºF to 104ºF)
Atmospheric pressure: 800 to 1060hPa
Environment - Transport & Storage
Relative humidity: 10% to 80%
Ambient temperature: -20ºC to 70ºC (-4ºF to
158ºF)
Atmospheric pressure: 500 to 1110hPa
kg
lbs
Safe Working Load
190
(420)
Maximum weight limit to be lifted or carried (when using ‘standing sling’)
190
(420)
Maximum weight limit to be lifted or carried (when using ‘transfer and
walking sling’ for walking practice only)
190
(420)
Maximum weight limit to be lifted or carried (when using the ‘transfer and
walking sling’ for transfer operation only)
140
(308)
Maximum weight limit to be carried (when using ‘toilet commode seat and frame’)
190
(420)