12 AH-MIN V3
08
/
10
/17
3. Tactile inspection of the buttons:
All buttons should be operated and response tested.
Depress each button on the handset. Each button should have a positive audible click that is
also felt. The function being pressed should also operate.
If the button:
does not have a click feel
does not have an audible click sound
does not immediately operate the function
feels soft offering no resistance
feels different to the rest of the buttons
travels further when depressed
Immediately quarantine the bed and report the damage to maintenance staff
or replace the Handset, otherwise contact an ArjoHuntleigh approved Technician.
The following table identifies some fault symptoms, their possible cause and
suggest some simple checks and corrective actions. If these actions fail to
solve the problem consult the ArjoHuntleigh Service Department on Freecall
1800 072 040.
Symptom
Possible Cause
Action
Actuators fail to work
Power disconnected or
batteries discharged
(battery backup models
only)
Ensure mains supply lead is
connected
Handset disconnected
Check the handset plug is
fully inserted in the control
box
Blown fuse
Contact ArjoHuntleigh
Service Department on
1800 072 040
Damaged handset cable Replace handset
One actuator does not work Actuator plug
disconnected
Check the actuator plug us
fully inserted in control box.