22
Care and Preventive Maintenance
The
Maxi Air
system is subject to wear and tear, and the following actions must be performed to make sure that the
product remains within its original manufacturing specification.
Preventive maintenance schedule for
Maxi Air
air supply
and the
Maxi Air
mattress
.
WARNING
To avoid malfunction resulting in injury, make sure to conduct regular inspections and follow
the recommended maintenance schedule. In some cases due to heavy use of the product and
exposure to aggressive environment more frequent inspections should be carried out. Local
regulations and standards may be more stringent than the recommended maintenance
schedule.
WARNING
To avoid injury to both patient and caregiver, never modify the equipment or use incompatible
parts.
CAREGIVER OBLIGATIONS Action/Check
Before
Every New
Patient
Every
Transfer
Every
Week
Every
Month
Disinfect (only the
Maxi Air
air supply)
X
Visually check all exposed parts
X
Perform functionality test
X
Check/change the air filter
X
WARNING
To avoid injury and/or unsafe product, the maintenance activities must be carried out at the
correct frequency by qualified personnel using correct tools, parts and knowledge of proce-
dure. Qualified personnel must have documented training in maintenance of this device.
QUALIFIED PERSONNEL Action/Check
Every
Year
Change the air filter
X
Perform functionality test
X
NOTE
All Caregiver Obligations are to be checked when performing the Qualified Personnel Service.
Summary of Contents for Maxi Air
Page 1: ... with people in mind 04 MS 01_7GB July 2013 Maxi Air INSTRUCTIONS FOR USE ...
Page 9: ...9 Intentionally left blank ...
Page 11: ...11 14 17 11 10 16 18 9 15 12 13 1 2 3 4 5 6 7 8 ...
Page 15: ...15 Intentionally left blank ...
Page 29: ...29 Intentionally left blank ...
Page 31: ...31 Data label LATEX REF Maxi Air MAS014000 WW LOT Patient Name Date L ...