45
Resolution
0.1 °C (0.18 °F)
NOTE
All drainage, water and electrical installation work should be carried out by qualified personnel, in
accordance with local and national regulations.
Operating Conditions
Ambient temperature
10 °C – 40 °C (50 – 104 °F)
Relative humidity
30% - 95% Rh
Atmospheric pressure
800 hPa to 1060 hPa
Transport and Storage
Ambient temperature
-20 °C to 70 °C (-4 °F to 158 °F)
Relative humidity
10% - 80% at 20 °C (68 °F)
Atmospheric pressure
500 hPa to 1100 hPa
With Height Adjustable Bathtub and/or Lift Arm
Voltage
230 V single phase AC
Frequency
50 Hz
Power consumption
800 VA
Fuses
230 V 10 A
Power supply cord
Only replaced by qualified personnel.
Lifting stroke (bathtub)
300 mm (11 3/4”)
Bathtub raising time (full load)
33 ± 5 s
Bathtub lowering time (full load)
25 ± 5 s
Lifting stroke (lift arm)
Malibu/Sovereign
bath
530 mm (20 7/8”) / 580 mm (23”)
Cyclic time for lift arm movement
(up-turn-down, full load)
65 ± 5 s
Maximum duty cycle: Lift actuator
1 minute ON
9 minutes OFF
Maximum duty cycle: Air Spa system
15 minutes ON
15 minutes OFF
WARNING
To avoid electric shock, make sure that the equipment is connected to:
• Continuously powered supply mains with protective earth.
• Separate fuse and ground fault circuit interrupter (GFCI)
• A mains disconnection device
• Equipotential bonding point
All installations must be in accordance with local codes and regulations.
General
Summary of Contents for Malibu
Page 1: ... with people in mind 04 AZ 00_14GB 12 2018 Malibu Sovereign Instructions for Use ...
Page 47: ...47 Intentionally left blank ...
Page 51: ...51 Intentionally left blank ...
Page 56: ...Intentionally left blank ...
Page 57: ...Intentionally left blank ...
Page 58: ...Intentionally left blank ...