30
Always Follow these 21 Steps for Proper
Cleaning and Disinfection.
If an ArjoHuntleigh shower panel will be used, see respective
IFU
.
Removing visual residue (Step 1-3 of 21)
1
Detach pillow, cushions and safety belt.
2
Clean the
Carino
from visible residue using a cloth soaked
in water or if more suitable by spraying/showering water
using a brush and/or wipe with a clean cloth. Start from the
top and move downwards.
3
Clean all removed parts from visible residue with a cloth
soaked in water or shower water using a brush and/or wipe
with a clean cloth (if applicable).
Cleaning (Step 4-10 of 21)
4
Put on protective gloves and eye wear.
(See Fig. 1)
5
Spray disinfectant on the
Carino
. Use a brush and scrub
thoroughly, especially on handles and hand control.
(See Fig. 2-3)
6
Rinse the
Carino
thoroughly with water (approx. 25 °C)
until all the disinfectant has been removed.
7
Spray disinfectant on the removed parts. Use a brush soaked
in disinfectant for cleaning, scrub thoroughly.
8
Rinse the removed parts thoroughly with water until all the
disinfectant has been removed.
9
Soak a cloth in disinfectant and wipe all other areas such as
chassis.
10 Wet a new cloth with water and remove all traces of
disinfectant on other areas. The cloth must occasionally be
rinsed in running water when removing disinfectant.
continue with the steps on the next page
Fig. 1
Fig. 2
Fig. 3
Summary of Contents for Carino
Page 1: ... with people in mind Carino 04 BO 01_12US CA September 2014 INSTRUCTIONS FOR USE ...
Page 9: ...9 6 1 11 10 7 9 8 5 2 3 22 23 19 13 24 20 21 12 18 16 13 17 14 15 13 25 4 ...
Page 15: ...15 Intentionally left blank ...
Page 37: ...37 Intentionally left blank ...
Page 41: ...41 Ø1082 42 5 8 Ø1370 53 7 8 630 24 3 4 424 16 5 8 624 24 5 8 475 18 3 4 ...