9
Electromagnetic Compatibility
001-24257-EN rev
. 4
Electromagnetic Immunity
Guidance and Manufacturer’s Declaration -
Electromagnetic Immunity - For all Equipment and Systems
The Wall Mounted Battery Charger is intended for use in electromagnetic environment
specified below. The customer or the user of the Wall Mounted Battery Charger should assure
that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic environment -
guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient / burst
IEC 61000-4-4
±2 kV, AC Mains
100 kHz repetition
frequency
±2 kV, AC Mains
100 kHz
repetition
frequency
Mains power quality should be that of
a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±2 kV AC Mains,
Lines to Ground
±1 kV, AC Mains,
Line to Line
±2 kV AC Mains,
Lines to Ground
±1 kV, AC Mains,
Line to Line
Mains power quality should be that of
a typical commercial or hospital
environment.
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
0%
U
T
; 0,5 cycle
At 0°, 45°, 90°,
135°, 180°, 225°,
270° and 315°
0%
U
T
; 1 cycle
and
70%
U
T
; 25/30
cycles
Single phase: at 0°
0%
U
T
; 250/300
cycle
0%
U
T
; 0,5 cycle
At 0°, 45°, 90°,
135°, 180°, 225°,
270° and 315°
0%
U
T
; 1 cycle
and
70%
U
T
; 25/30
cycles
Single phase: at 0°
0%
U
T
; 250/300
cycle
Mains power quality should be that of
a typical commercial or hospital
environment.
Power
frequency (50 /
60 Hz)
magnetic field
IEC 61000-4-8
30 A / m
50/60 Hz
30 A / m
50/60 Hz
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercials or hospital environment.
NOTE:
U
T
is the AC mains voltage prior to application of the test level.