Safety Instructions

Battery and Battery Charger 

Safety Practices

• The charger must be placed in a low traffic area

to prevent it from being hit.

• When  plugging / unplugging  the  charger’s

power cord, grip the plug itself. Do not pull on the

cord.

• Attach  any  loose  power  cord  to  prevent  cable

wear and to avoid strangulation.

• Install  the  charger  away  from  heat  sources  or

direct sunlight.

• The  charger  has  been  made  for  dry

surroundings.  It  must  not  be  installed  in

bathrooms, or other humid areas.

• Only use the charger in well ventilated areas.
• The charger must not be covered or exposed to

dust or water. The metal contacts must be kept

clean. 

• The charger is for use only with Arjo 24V lead-

acid battery type, as supplied with the lift.

• The battery pack may remain connected to the

charger  unit  when  it  is  fully  charged  without

damage.

• Do not short-circuit the battery pack. 
• Do  not  crush,  puncture,  open,  dismantle  or

otherwise  mechanically  interfere  with  the

batteries. 

• The power socket must be easily accessible. If

a faulty condition occurs, switch off the power by

removing the plug from the wall socket.

• Do  not  store  battery  packs  at  a  temperature

higher than 40

°

C (104 

°

F). Should the battery

pack casing crack and cause contents to come

in  contact  with  skin  or  clothing,  rinse

immediately  with  water.  If  content  comes  in

contact with eyes, rinse immediately with plenty

of water and seek medical attention.

• The “Pb” symbol appears next to the recycling

and trash bin symbols on the battery pack label.

This means that the battery contains lead and

that it must not be disposed of, but recycled.

General Information

Symbols Used

Symbols

Key to Symbols

This  symbol  points  out  the

manufacturer’s name and address.

May  also  point  out  manufacturing

date.

CE marking indicating conformity with

European  Community  harmonised

legislation.

This  symbol  means  that  the

product must be kept dry.

This  symbol  points  out  the

manufacturer’s catalogue number. 
This  symbol  points  out  the

manufacturer’s lot number. 

This symbol  means  that you must

refer to the 

Instructions for Use

.

Waste  Electrical  and  Electronic

Equipment  (WEEE)  -  do  not

dispose  of  this  product  in  general

household or commercial waste.

The device is meant to be operated

indoors only.

Symbols

Key to Symbols

Class II electrical equipment. 

This  term  refers  to  electrical

equipment in which protection against

electric shock does not rely on basic

insulation only. 
This  symbol  means  the  product  is

certified  by  the  Slovenian  Institute  of

Quality and Metrology.

This  symbol  means  the  product  is

certified by TUV Rheinland. 

This  symbol  means  that  product

complies  with  Product  Safety

Appliance & Material Standards.
This  symbol  means  that  the  product

has been approved by the Canadian

Standards Association.
This  symbol  means  that  the  product

complies  with  European  Norms

Electrical Certification.
RCM Australian  Compliance  symbol,

indicating compliance with applicable

electrical  safety,  EMC,  EME  and

telecommunications requirements.

Indicates  the  product  is  a  Medical

Device  according  to  EU  Medical

Device Regulation 2017/745.

SI

E

P S 

Summary of Contents for NDA8200

Page 1: ...INSTRUCTIONS FOR USE Wall Mounted Battery charger 001 24257 EN rev 4 08 2019 ...

Page 2: ... 5 How to Use the Charger 5 Charging your Battery 5 Troubleshooting 6 Technical Specifications 6 Technical Data 6 Dimensions 6 Electromagnetic Compatibility 8 Electromagnetic Compliance 8 Electromagnetic Immunity 9 Design Policy and Copyright and are trademarks belonging to the Arjo group of companies Arjo 2019 As our policy is one of continuous improvement we reserve the right to modify designs w...

Page 3: ...ximize the long term reliability safety and value of the product Contact your local agent for product replacement This product does not require maintenance and have no parts that can be replaced You can get additional copies of this manual from your local agent When ordering include the Instructions for Use product number see front page and product part number Manufacturer Information This product...

Page 4: ... appears next to the recycling and trash bin symbols on the battery pack label This means that the battery contains lead and that it must not be disposed of but recycled General Information Symbols Used Symbols Key to Symbols This symbol points out the manufacturer s name and address May also point out manufacturing date CE marking indicating conformity with European Community harmonised legislati...

Page 5: ...on proper charging practices Recharge the batteries on a regular basis to prolong their life span Avoid the batteries from reaching a low charge state Remove the battery pack from the lift when it is not used for a long period Charging your Battery 1 Insert the battery pack into the charger unit and push it in firmly see Fig 2 2 Connect the charger into an AC outlet using the proper power cord 3 T...

Page 6: ...iations The Wall Mounted Battery Charger is also compliant to EN 60335 1 2012 A11 2014 and EN 55014 1 2006 A1 2009 WARNING Portable RF communications equipment including peripherals such as antenna cables and external antennas should be used no closer than 30 cm to any part of the Wall Mounted Battery Charger including cables specified by the manufacturer Otherwise performance degradation of this ...

Page 7: ...Technical Specifications 7 001 24257 EN rev 4 Dimensions Fig 4 715mm 28 1 8 100mm 3 15 16 60mm 2 3 8 305mm 12 46mm 1 13 16 45mm 1 3 4 PRODUCT IDENTIFICATION LABEL CHARGER STATUS INDICATOR LED ...

Page 8: ...eration of nearby equipment It may be necessary to take action such as reorienting relocating the equipment or shielding the location WARNING Use of accessories cables and spare parts other than those specified or provided by Arjo could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation Guidance and Manufacturer s ...

Page 9: ...ency Mains power quality should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 2 kV AC Mains Lines to Ground 1 kV AC Mains Line to Line 2 kV AC Mains Lines to Ground 1 kV AC Mains Line to Line Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4...

Page 10: ...een 0 15 80 MHz N A Radiated RF IEC 61000 4 3 10 V m 80 MHz to 2 7 GHz 10 V m 80 MHz to 2 7 GHz N A Proximity fields from RF wireless communications equipment IEC 61000 4 3 380 390 MHz 27 V m PM 50 18 Hz 430 470 MHz 28 V m FM 5 kHz 1 kHz sine PM 18 Hz 800 960 MHz 28 V m PM 50 18 Hz 1700 1990 MHz 28 V m PM 50 217 Hz 2400 2570 MHz 28 V m PM 50 217 Hz 5100 5800 MHz 9 V m PM 50 217 Hz 704 787 MHz 9 V ...

Page 11: ... D 7 UMR WDOLD 6 S 9LD LDFRPR 3HURQL 7 520 7HO D PDLO WDO SURPR DUMR FRP 0 67 UMR 0LGGOH DVW 2IILFH WK ORRU 4 XLOGLQJ 1RUWK 7RZHU XEDL 6FLHQFH 3DUN O DUVKD 6RXWK 3 2 R XEDL 8QLWHG UDE PLUDWHV LUHFW D PDLO QIR 0 DUMR FRP 1 5 1 UMR 9 LH HQZHL 0 7 3RVWEXV 7 7HO D PDLO LQIR QO DUMR FRP 1 1 UMR WG 9HVWH ULYH 0RXQW HOOLQJWRQ 1 8 1 7HO UHH DOO D PDLO Q LQIR UMR FRP 125 UMR 1RUZD 6 2ODI HOVHWV YHL 1 26 2 ...

Page 12: ...re ergonomic patient handling personal hygiene disinfection diagnostics and the effective prevention of pressure ulcers and venous thromboembolism we help professionals across care environments to continually raise WKH VWDQGDUG RI VDIH DQG GLJQL HG FDUH YHU WKLQJ ZH GR ZH GR ZLWK SHRSOH LQ PLQG ArjoHuntleigh AB Hans Michelsensgatan 10 211 20 Malmö Sweden www arjo com ...