43
Troubleshooting
WARNING
To avoid injury, make sure that the patient is not left unattended at any time.
PROBLEM
ACTION
Electrical movement of the
Carevo
shower
trolley and adjustable back support (if available)
continues after releasing the control buttons.
1 Activate the emergency stop button.
2 Remove the patient from the
Carevo
shower trolley and
contact qualified personnel.
Electrical movement of the
Carevo
shower
trolley and adjustable back support does not work
when pressing the control buttons.
1 Make sure that the
Carevo
shower trolley movements
are not blocked (by a bed or something nearby the
Carevo
shower trolley).
2 Press the button on a different control panel. If that
works, contact qualified personnel to replace the
damaged control panel.
3 Make sure that the emergency stop button is inactivated.
To inactivate turn the emergency stop button clockwise.
4 Check that the battery is correctly inserted in the battery
compartment.
5 Make sure the battery is charged.
6 Contact qualified personnel.
The opening handles will not lock.
1 Remove the patient and make a visual check.
2 Contact qualified personnel.
The
Carevo
shower trolley is hard to manoeuvre
during transport.
1 Make sure that all the brakes and straight steering
devices are released.
2 Remove the patient from the
Carevo
shower trolley
and
check that all the castors swivel and roll freely.
3 Contact qualified personnel.
It is only possible to lower the lift.
Lower devise to a safe position. Stop using the equipment and
contact authorized service personnel.
NOTE
If the problem can not be solved with the written troubleshooting actions, please contact qualified
personnel.
Summary of Contents for Carevo
Page 1: ...04 BA 08_13EN 06 2019 Carevo 04 BA 08_13EN 06 2019 INSTRUCTIONS FOR USE ...
Page 11: ...11 3 8 7 9 23 15 1 2 6 12 21 8 18 25 17 19 4 5 24 10 22 2 20 11 13 16 14 26 ...
Page 31: ...31 Intentionally left blank ...
Page 33: ...33 Fig 1 ...
Page 37: ...37 Intentionally left blank ...
Page 51: ...51 Intentionally left blank ...