Care and Preventive Maintenance

WARNING

To avoid injury to both patient and 

caregiver, never modify the equipment or 

use incompatible parts.

Before and After Every Use

Visually check all exposed parts

The caregiver shall inspect the sling before and after 

every use. The complete sling should be checked for 

all deviations listed below. If any of these deviations 

are visible, replace the sling immediately.

• 

Fraying

•  Loose stitching

• 

Tears

•  Holes

•  Discolouration and stains from bleaching

•  Sling soiled or stained 

• 

Unreadable or damaged label

• 

For 

AHD001

: Make sure the “Do NOT wash” 

symbol on the head support is intact. If damaged or 

missing and the underlying red “Do NOT use/lift” 

symbol appears, replace the sling immediately. See 

“Label on the Sling” on page 16.

When Soiled, Stained and Between 

Patients

MAA6000

The caregiver should make sure that the sling is 

cleaned according to 

“Cleaning and Disinfection” 

on page 13

, when it is soiled or stained and between 

patients.

AHD001

Washing or disinfection is not allowed. Do not wipe, 

sterilize or soil the 

AHD001

. If the 

AHD001

 has been 

subjected to any treatment, it shall be discarded.

Storage

When not in use, the slings should be stored away 

from direct sunlight where they are not subject to 

unnecessary strain, stress or pressure, or to excessive 

heat or humidity. The slings should be kept away from 

sharp edges, corrosives or other things that could cause 

damage on the sling.

Service and Maintenance

It is recommended that qualified personnel inspect the 

condition of the sling twice a year (every 6 month), 

according to ISO 10535.

Troubleshooting

Problem

Action

The patient is not positioned correctly in 

the sling.

•  Make sure that all loops are attached at the right 

length.

•  Make sure the straps are not twisted.

•  Make sure the patient is positioned on the inside of 

the sling. The outside has a label for reference.

•  Make sure the patient is centred in the sling.

The patient feels discomfort when using 

the sling.

•  Make sure that there are no creases on the fabric.

•  Make sure that all loops are attached at the right 

length.

It is difficult to use the sling in combination 

with the lift.

Use the sling only in accordance with the combinations 

described in section 

“Allowed Combinations” 

on page 7

.

The position of the patient in the sling 

should be more reclined or upward.

Use the loop lengths to adjust the patient’s position.

The patient wishes to have extra support 

for the neck.

Place a pillow behind the patient’s neck, inside the sling.

AHD001 is soiled, stained or wet.

Dispose AHD001 and replace it with a new one.

The “Do NOT wash” label is damaged or 

missing and the underlying red “Do NOT 

use/lift” label appears on the AHD001

Dispose AHD001 and replace it with a new one.

Summary of Contents for AHD001

Page 1: ...INSTRUCTIONS FOR USE Repositioning Slings 04 SQ 00 EN_7 11 2021...

Page 2: ...the Instructions for Use Design Policy and Copyright and are trademarks belonging to the Arjo group of companies Arjo 2021 As our policy is one of continuous improvement we reserve the right to modify...

Page 3: ...sident Assessment 4 Expected Service Life 4 Safety Instructions 5 Serious incident 5 Parts Designation 5 Preparations 6 Before First Use 6 steps 6 Before Every Use 7 steps 6 After every use 6 Sling Se...

Page 4: ...e patient resident shall be passive Might be almost completely bed ridden Often stiff contracted joints Totally dependent on the caregiver Physically demanding for caregiver Stimulation and activation...

Page 5: ...s being used WARNING To avoid injury never let the patient smoke when using the sling The sling can catch on fire WARNING To avoid injury only store the equipment for a short period of time If stored...

Page 6: ...ling Check for fraying loose stitching thread breakage tears fabric holes soiled fabric damaged loops unreadable or damaged label WARNING To prevent cross contamination always follow the disinfection...

Page 7: ...lbs One Size One Size One Size AHD001 272 kg 600 lbs One Size One Size One Size Lift Maxi Move SWL 125 kg 275 lbs 227 kg 500 lbs Spreader Bar 2 point loop Small 2 point loop Medium 2 point loop Sling...

Page 8: ...centre line outside facing the outside The sling label is located on the outside WARNING To avoid injury during log roll make sure there is another caregiver present and or use the side rails on the o...

Page 9: ...Make sure that all loops are attached securely and the patient lays comfortably in the sling Fig 6 Fig 7 Fig 8 Fig 9 Turning to the Side Follow these 9 steps 1 Position the spreader bar in line with...

Page 10: ...lays comfortably in the sling 6 If an adjustment is needed lower the patient and make sure that the weight of the patient is taken up by the receiving surface before removing the loops 7 If possible...

Page 11: ...ig 15 5 Lower the spreader bar to enable the straps to be attached Caregiver should always hold on to the spreader bar to avoid swinging movements See Fig 16 6 Attach the loops Use the loop lengths th...

Page 12: ...e Repositioning Sling make sure the Repositioning Sling is not attached to any other object than the spreader bar make sure that there is clearance to perform the manoeuvre 11 If an adjustment shiftin...

Page 13: ...g to local hygiene legislation If allowed according to the product label tumble dry in low temperature max 60 C 140 F 4 Do NOT wash with other items that have rough surfaces or sharp objects use any m...

Page 14: ...tress or pressure or to excessive heat or humidity The slings should be kept away from sharp edges corrosives or other things that could cause damage on the sling Service and Maintenance It is recomme...

Page 15: ...See Sling Selection on page 6 Operating Transport and Storage Environment Temperature 0 C to 40 C 32 F to 104 F Operating and Storage 25 C to 60 C 13 F to 140 F Transport Humidity 15 70 at 20 C 68 F...

Page 16: ...ulation 2017 745 AHD001 Symbols Do NOT wash symbol Located on the outside of the sling Do NOT use lift symbol The AHD001 has been washed Located on the outside of the sling Symbol for Loop Use a loop...

Page 17: ...Intentionally left blank...

Page 18: ...Intentionally left blank...

Page 19: ...oor HQ Building North Tower Dubai Science Park Al Barsha South P O Box 11488 Dubai United Arab Emirates Direct 971 487 48053 Fax 971 487 48072 Email Info ME arjo com NEDERLAND Arjo Nederland BV Biezen...

Page 20: ...ensure ergonomic patient handling personal hygiene disinfection diagnostics and the effective prevention of pressure ulcers and venous thromboembolism we help professionals across care environments to...

Reviews: