HYDROSENSE HS2410
Revision 5.0
P a g e 12 | 27
3.3 UNIT
INSTALLATION
1)
Locate an area that is environmentally protected from wide variances in temperature and
humidity. Indoor installations are recommended.
When selecting the location, consider that regular maintenance and testing is desirable for the
proper and accurate operation of the instrument.
If the sample input is not from a pumped source, locate the instrument in a position that will
receive a continuous representative sample from the process stream. The farther from the
stream, the greater the lag time of readings vs. actual process variance.
2)
Mount the enclosure on a vertical wall or support of minimal vibration. The sample will be
flowing over a Flow Plate. A bubble level is provided in the sample chamber. Be sure the unit is
mounted level. This is necessary for a proper flow across the Flow Plate.
3)
Open the Blue Lamp Box and secure the new lamp into place with the metal band and knurled
screw. Make sure the lamp is seated between two tabs of metal band as per picture below.
Figure 1 – SECURE LAMP INSIDE THE LAMP BOX
4)
Plug the lamp terminal into the socket. Close the Blue Lamp Box.
TABS
SEATED BETWEEN TABS
KNURLED
SCREWS
METAL
BAND