HYDROSENSE HS2410
Revision 5.0
P a g e 22 | 27
7.5.2 FILTER
<2>
Press <2> for filter. The moving average filter tracks the average of the last # of samples. Higher values
provide more stable readings. Max value is 1000. The default value is 960.
7.5.3 UNITS
<3>
Press <3> for units. There are 3 units are available: ppm, FLR and mg/L. Press <1> for ppm, <2> for
FLR and <3> for mg/L.
7.5.4 PASSWORD
<4>
Press <4> for password. Enter the new password at the prompt, then press <ENTER>.
7.5.5 MODBUS ADDRESS <5>
Press <5> for Modbus address. The Modbus address is used only for network applications and is
usually factory set. To communicate on a network, each controller must have a unique Modbus
address.
Important:
IF MULTIPLE UNITS ON A NETWORK HAVE THE SAME ADDRESS, NETWORK ERROR
WILL RESULT. ENTER THE DESIRED TAG NUMBER.
Enter the desire modbus address, press <ENTER> to Settings Menu.