Figure 15 – Calibration Reminder Screen View
This unit allows for three types of calibration.
7.8.1 Quick Calibration
This would be the most typical calibration and uses a factory defaulted slope to allow for a single
entry calibration point.
Figure 16 – Quick Calibration Set-up Screen View
Display Units
Choose the engineering units you want on the display, such as inches, mm, cm, liters, etc. (Factory
default is mm)
After entering each value for the following parameters, Press, “Enter” to confirm. Screen will change
to “Entered” if accepted.
Depth of oil for 100%
For the chosen engineering units, enter the total oil thickness that will represent 100% oil. This will
typically be 25mm for sensor A00761 and 300mm for sensor A00765.
Present Depth of surface oil
Enter the present thickness of surface oil on the water. Enter this in your chosen display units. If no
oil is present, enter 0.0.
Slope
This is factory preset. The factory default is in pF/mm. If the display engineering units have been
changed (ie. to inches), the slope must be mathematically corrected and re-entered.