AZI Customer Service 800-528-7411 or 602-470-1414
Page 22 of 57
Daily Operations
Before each day's use of the Jerome
®
J605, perform the following steps to verify proper
instrument operation:
•
Press the power
I/O
button to turn the instrument on.
The display will light up and show instrument serial number and software revision.
If necessary, press
ESC
to clear any calibration reminders. Call AZI Customer
Service at 800-528-7411 or 602-470-1414, or e-mail
, to schedule
instrument calibration.
The digital meter displays 0.000 ppm (or 0.00 ppb, depending on what Range is
currently selected).
Check the battery level as indicated by the battery icon at the top center of the
instrument display.
If the battery meter is empty and flashing, refer to
If the battery meter is not empty, but is flashing, then the instrument is currently
charging the battery.
When the instrument is plugged in and powered off, the display will stay active
and indicate “Charging.”
To ensure the instrument's electronics have stabilized, allow a 5-minute warm up
before beginning the next step.
•
Perform sensor regeneration. Refer to
•
Ensure the instrument has been powered on for at least five (5) minutes prior to sampling.
•
Use the Zero Air Filter to equilibrate the instrument to ambient air temperature.
Install the Zero Air Filter in the instrument’s intake.
Sample repeatedly every 15 seconds until the readings stabilize, then remove the Zero
Air Filter.
•
Press the SAMPLE button (at the end of the handle).
•
At the end of the sampling cycle, read the digital meter.
The number shown on the digital meter is the hydrogen sulfide concentration in ppb or
ppm as appropriate.
This sampled hydrogen sulfide value remains on the display until the next sample is
taken.
The digital meter automatically re-zeroes at the start of each sample.
•
At the end of each day's use, perform sensor regeneration as described in the next section.
DO NOT ALLOW HYDROGEN SULFIDE TO STAY
ON THE GOLD FILM SENSOR OVERNIGHT.
NOTE:
For levels less than 10 ppb, it is necessary to run a warm-up routine before sampling.
To initiate the automatic five minute warm-up, install a
Zero Air Filter (AZI P/N Z2600 3905)
in the intake, and select
Warmup
For levels of 10 ppb and above, the warm-up routine is not necessary.