UM Agitatore magnetico riscaldante digitale M2-D Pro ENG rev.0 11.05.2020
If you need an external temperature probe, it is advisable to connect it before turning on the instrument.
Procedure:
•
Make sure that any external probe is well connected to the instrument.
•
Set the temperature using the adjustment knob and bring it to the desired value.
•
When the heating function is active, the “Heat” LED lights up and the display indicates the real temperature.
•
The set temperature will be displayed on the right side of the screen.
•
The heating function is turned on / off by pressing the temperature
adjustment knob.
Operation with external temperature probe inserted
The word "Probe" will appear permanently on the display to indicate that the sensor
is operating.
Basic stirring
The stirring function is activated or deactivated by pressing the stirring speed adjustment knob. The speed can be set
between 100 and 1500 rpm in 10 rpm steps. When both functions are active and all speed adjustment operations have
been performed, the display returns to the speed and temperature display in about 5 seconds.
Remote control
The instrument can be controlled from an external PC (through
the appropriate software) through the RS232 serial port located
on the back of the instrument.
The possible transmission of data to the PC is possible only at the
request of the PC itself.
•
Transmission method: Asynchronous signal
•
Transmission mode: Fully Duplex.
•
Transmission speed: 9600 bit / s
Malfunctions
• In case of instrument failure, it is recommended to turn it off;
• Turn off the main ON / OFF switch for a few seconds and turn the unit back on;
• The stirring function will
continue to operate at the set speed before the fault occurs;
• The heating function will continue to operate at the set temperature before the fault occurred;
• If the problem is not resolved, contact technical support.
PT1000 connection
RS232
Power