UM Agitatore magnetico riscaldante digitale M2-D Pro ENG rev.0 11.05.2020
Control
Description
Temp
e
°C
These icons appear when the function heating is active.
Stir
This icon appears when the stirring function is active.
Mode
View the working mode.
Hot
Warns of the possibility that the plate is still too much hot
and therefore potentially dangerous. This function is also
active when the function is deactivated heating if the
plate temperature is above 50 ° C.
Probe
The external probe is connected to the instrument.
Remote
The instrument is controlled remotely.
Err
Displays an operating error.
Valore reale e valore impostato
Display of the heating and / or stirring parameters in
progress.
Note:
If the heating and stirring function were started at the same time, the display of the "heating" function always
has priority. If the speed is changed, the new value is displayed for 5 seconds.
Operating modes
Operating mode A
The user can adjust the parameters via the front controls or via a PC connected to the RS232 port.
The residual temperature safety function, the safety temperature limit and the automatic braking function for detaching
the magnetic anchor, can be changed.
Operating mode B
This mode is active only with an external probe connected.
The user can adjust the parameters via the front controls or via a PC connected to the RS232 port.
The residual temperature safety function, the safety temperature limit and the automatic braking function for detaching
the magnetic anchor, can be changed.
Operating mode C
The instrument keeps in memory the last speed and temperature settings set before the last shutdown.
PC remote control can be used but without the connection of the external probe.
The residual temperature safety function, the safety temperature limit and the automatic braking function for detaching
the magnetic anchor, CANNOT be changed.
Heating function
The device control system is digital and the temperature control system in particular has two separate safety circuits.
The plate is kept at a constant temperature by a digital control circuit. The plate temperature is monitored by two
internal temperature sensors (Pt1000) integrated in the plate itself. The external Pt1000 probe is able to monitor the
temperature of the heated sample.