Control 

Description 

Temp 

°C

 

These icons appear when the function heating is active. 

Stir 

This icon appears when the stirring function is active. 

Mode 

View the working mode. 

Hot 

Warns of the possibility that the plate is still too much hot 
and therefore potentially dangerous. This function is also 
active  when  the  function  is  deactivated  heating  if  the 
plate temperature is above 50 ° C. 

Probe 

The external probe is connected to the instrument. 

Remote 

The instrument is controlled remotely. 

Err 

Displays an operating error. 

Valore reale e valore impostato

 

Display  of  the  heating  and  /  or  stirring  parameters  in 
progress. 

 
Note: 

If the heating and stirring function were started at the same time, the display of the "heating" function always 

has priority. If the speed is changed, the new value is displayed for 5 seconds.

 

 

 

Operating modes 
 

Operating mode A 

The user can adjust the parameters via the front controls or via a PC connected to the RS232 port. 
The residual temperature safety function, the safety temperature limit and the automatic braking function for detaching 
the magnetic anchor, can be changed. 
  

Operating mode B 

This mode is active only with an external probe connected. 
The user can adjust the parameters via the front controls or via a PC connected to the RS232 port. 
The residual temperature safety function, the safety temperature limit and the automatic braking function for detaching 
the magnetic anchor, can be changed. 

 

Operating mode C 

The instrument keeps in memory the last speed and temperature settings set before the last shutdown. 
PC remote control can be used but without the connection of the external probe. 
The residual temperature safety function, the safety temperature limit and the automatic braking function for detaching 
the magnetic anchor, CANNOT be changed. 

 

 

Heating function 

The device control system is digital and the temperature control system in particular has two separate safety circuits. 
The  plate  is  kept  at  a  constant  temperature  by  a  digital  control  circuit.  The  plate  temperature  is  monitored  by  two 
internal temperature sensors (Pt1000) integrated in the plate itself. The external Pt1000 probe is able to monitor the 
temperature of the heated sample. 
 
 