Page 9
5.4. Entering of the Cut Off « CUTOFF »
CUT U / CUT I
05 . 0
Programming of the voltage and current cut offs (value below which the instrument will
%
enforce the measure to 0). To be entered in % of the caliber value (1 or 5A, 100 or
330V).
5.5. Processing « FONCT »
Choice of the digital fitlering « FILTER »
OFF
Choice of the digital filtering weight.
NR 1
Increase the value in case of unsteady measures.
NR 2
NR 3
De-validation of the errors « DIAG »
OFF
Allows validating or de-validating the measure errors.
ON