X. DETERMINATION OF DISTANCE TO PATIENT’S TEETH
XI. PROBLEMS AND SOLUTIONS
The device does not work.
Check if the power cord from the adapter is plugged into the power network and all cable connectors are
correctly coupled.
Damaged power cord.
Do not use the device! Switch it off and immediately take it to the service facility
Unpleasant or irritating thermal sensation by the patient.
Increase the distance of the device to the patient’s teeth to 10-15 cm. If the thermal sensation is still unpleasant
at this distance switch to low intensity- “Lo” (Low) and shorten the distance to the teeth.
The display is blinking and no light is emitted.
The device is in PAUSE mode. To resume operation, press either + or - buttons on Bleaching Head.
The Bleaching Head cannot be fixed.
Read Point 5. In Section V. INSTALLATION and slightly adjust the tension of the Bleaching Head until the desir-
able mobility is reached.
The display only shows “Oh”.
In case of overheating the thermal protection is triggered and the device will cease to emit light. The display
will indicate “Oh” and the fans will turn on for 1 minute. After cooling, the device will continue its normal op-
eration. Overheating can only happen in an emergency situation or by accidental clogging of the cooling
vents. That’s why the cooling vents located on the bottom of the bleaching head must always be unob-
structed in order to allow air circulation.
For all other questions regarding the installation and/or operation of the device, please contact the manu-
facturer or your local authorized representative.
5 cm distance
5 cm distance
9 cm distance
9 cm distance
15 cm distance
15 cm distance
HIGH intensity
LOW intensity
Hot
Accetable
Warm
Weak
TOUCH
LIPS
CUBE
ULTRA
CLICK
STICK
DUETTO
DUETTO
ARCUS
HOME
DAY
ACTIVE
HOME
NIGHT
HOME
ARCUS
CARE
BRUSH
SEAL
DUETTO
CARE
HOME
FAST
HOME
MILD
HOME
FAST
HOME
MILD
LED Bleaching Lamp - FIXED
13
GB
Summary of Contents for BLANCONE
Page 91: ......