11

10

Only qualified/authorized engineers or service personnel should 

install the unit

. Installation by unqualified personnel may cause 

electric shock or fire.

Always make sure the unit is grounded to protect the user from 

electrical shock.

Do not use the power supply cord if it is damagaged.

 Such supply 

cord may cause fire or electric.

Never use a telephone line or lightning rod as ground protection. 

During lightening, there is a strong current present, which is 

extremely dangerous.

Electrical connection

In order to ensure a reliable installation that complies with the limits of 

temperature and voltage drop, it is necessary to determine maximum load of 

appliances connected to the installation. In determining the maximum load 

for an installation or for a part of it, it is vital to take into account 

contemporaneous factors. For supply systems, the following must be 

determined:

- Power system (AC/DC) - Data regarding voltage and absorbed power/

  current is given on the rating plate. 

- Ground protection: To prevent the user of getting exposed to electrical 

  shock, in the event of a insulation damage, the appliance must be grounded

- The installation must always be secured with a minimum 10 A fuse.

- If more than one appliance needs to be installed, each appliance must be 

  connected to an individual fuse group. 

 

When installing the unit, make sure the protected earth is grounded. If the 

connection is a 3-prong connection, use a three-pin plug and connect the 

conductor with yellow / green insulation to ground. In order to maintain a 

stable operation of the unit, voltage variation cannot differ more than ±10 per-

cent of the nominal voltage supply.  

Always follow local regulations when preparing an installation. If in doubt 

always contact your local authorities.

Do not use water pipes as ground protection. Modern water pipe 

systems are non-conductors such as PVC.

Replacing the power cord may only be done by authorized 

personal.

Never use gas lines as the ground protection for the unit. This can 

cause an explosion.

Disconnect the power cord if there is something wrong with the 

unit. 

Continued abnormal operation may cause electric shock or fire.

Getting started

During the initial startup and continuous usage of the appliance, the following 

procedures shall be followed.

1. Plug the power socket into a dedicated outlet. For correct voltage 

   requirement, follow the information on the rating plate. 

When started, let the appliance cool for 24 hours before placing any 

products into it.

2. When the appliance initially starts up, the buzzer alarm may sound. This is 

    normal. You can accept the alarm on the display.

3. This appliance has been set to operate at -80°C from the factory.

4. Check the functionality of the light by opening the door. 

5. Once a thorough inspection of the appliance is completed, products can 

    be loaded into the appliance. Products should be pre-frozen when inserted 

   into the appliance, otherwise this can affect the cooling performance of the  

   appliance.

Summary of Contents for ULUF P390

Page 1: ...ORIGINAL OPERATING INSTRUCTIONS Integraline RANGE ULUF P390 P500 P610 P820...

Page 2: ...vents Compressor failure alarm Outcome frames Contents Page 4 4 4 4 5 5 5 6 6 6 6 7 7 8 8 9 10 11 12 13 15 15 16 16 17 17 18 18 18 19 20 21 22 Warnings How to turn on the device Menu user How to chang...

Page 3: ...ctiko is obligated as a supplier to ensure the users safety when operating one of our appliances To prevent personal and material injury or damage please follow the instructions in this manual Intende...

Page 4: ...Be careful not to tip over the appliance during transport to prevent damage or injury Plastic bags pose a suffocation risk Keep away from children Inactivity for extended period If the appliance must...

Page 5: ...n or der to achieve optimal operating results Firm and levelled floor Placing the appliance on a firm and flat floor reduces the risk of excessive noise and vibration Away from any heat sources Avoid...

Page 6: ...connection is a 3 prong connection use a three pin plug and connect the conductor with yellow green insulation to ground In order to maintain a stable operation of the unit voltage variation cannot d...

Page 7: ...filter less construc tion for less maintenance low noise level low energy consumption low heat dissipation and excellent temperature stability The compressors and cooling system is located in the top...

Page 8: ...r openings and control the lights 15 Cover for noise reduction and air flow management 5 Probe See Probes and portholes 16 Remote alarm connection For connecton of additional alarm equipment 6 Door ha...

Page 9: ...ve keyboard You can access a variation of settings in order to adjust your unit for your exact needs A 4GB microSD is provided as standard installed internally for recording functional data every 30 s...

Page 10: ...19 18 Info events Door Warning Action Icons Temperature alarms...

Page 11: ...oller The other compressor will keep running To reset the alarm and reenable the compressor accept the alarm and turn the unit off and on again During pulldown this alarm may be activated If this happ...

Page 12: ...23 22 Outcome frames The outcome frames will be accompanied by a long beep Warnings To accept a warning press DOOR OPEN...

Page 13: ...press Enter Menu user STAND BY SETPOINT LIMITS ACCESS LIST ECOMODE EVENTS GRAPHICS DEFROST SETTINGS SERVICE press and hold to turn off the controller change the temperature setpoint set the high low...

Page 14: ...27 26 How to change the setpoint How to change the alarm limits Press the Menu key to access the menu...

Page 15: ...29 28 Access list Ecomode setting KEY FUNCTION...

Page 16: ...30 Event list If the event list is empty the icon above will be shown Shows number of pages with events Press or to change page Graphs SCROLLING THE FRAPH AND ENABLING THE TRACES REPRESENTATION CRITER...

Page 17: ...edit the telephone numbers where to send the alert messages set the language of user interface edit machine name by alphanumeric keyboard How to set date hour Press select to move the highlight and t...

Page 18: ...is 111 Service For accessing the service menu The service settings are reserved for qualified Arctiko service personnel By entering the admin password both the admin and user passwords can be changed...

Page 19: ...Pressing the backup icon shows the usb pen insertion request frame with 20 second time out If the usb pen has been recognised the setting page of the months to download is displayed otherwise the disp...

Page 20: ...stributor 2 To convert the data open the program Thermo Converter To convert the file press the DAT TO CSV CONVERT button 1 Insert the USB drive with the loaded data in a computer 3 Open the file by f...

Page 21: ...ght corner of the compressor department 1 Logo 2 Contact information 3 Model 4 Voltage range 5 Refrigerant and quantity 6 WEEE Waste from Electrical and Electronic Equipment 7 Compressor type 8 CE mar...

Page 22: ...operation 1h RELAIS U2 U2 Relay status COMP 24h Compressor operation 24h RELAIS U3 U3 Relay status COMP ON Time on compressor RELAIS U4 U4 Relay status COMP OFF Time o compressor RELAIS U5 U5 Relay st...

Page 23: ...ssor compartment using only a dry brush or a vacuum cleaner Parts in the refrigeration system for this appliance are completely sealed These do not require any lubrication If you would like more infor...

Page 24: ...unit away from the sunlight or heat source Frequent door openings may cause the temperature to rise Check if there has been frequent door openings Leave the door closed until the temperature in the un...

Page 25: ...5080063 02 GB Arctiko A S Oddesundcej 39 DK 6715 Esbjerg N Phone 45 7020 0328 info arctiko com arctiko com...