Defrosting 

This appliance does not have automatic defrost, therefore you have to 

perform defrosting by following these instructions:

  

Defrost the appliance at least twice a year. 

1. If the appliance is loaded with samples or products, move these to another    

  appliance to maintain the temperature of the products. 

2. Turn off the appliance on the display by pressing the STANDBY           

     button. 

  

3. Unplug the main power cable from the power socket. 

4. Open the door of the appliance. 

5. Place cloths or towels in the bottom of the appliance to collect any

    defrost water.

ATTENTION !

Do not use abrasive nor chlorine-containing products to clean the appliance.

Cleaning

Perform following instructions at least once a month:

Always keep the appliance free of ice.

 Use a soft cloth or brush to remove 

  loose ice. Never use sharp tools and be cautious not to damage the 

  gaskets. Keeping the appliance free of ice will extend its lifespan.

- Clean the outside and inside of the appliance using a dry, soft cloth or a soft 

  cloth with a solution of water and mild detergent. If a thorough cleaning or 

  disinfection is required, we recommend using ethanol.

Clean all gaskets

 using a damp cloth and if necessary a mild 

  detergent. Remove dirt and wipe with a dry cloth afterwards. Do not pour 

  water into the appliance. By doing so, the water can damage the insulation 

  materials and electrical components.

Dust off the inside of the compressor

 compartment using only a dry brush 

  or a vacuum cleaner. Parts in the refrigeration system for this appliance are 

  completely sealed. These do not require any lubrication.

If you would like more information about your device or you would like to 

purchase spare parts or additional equipment, please contact your local 

distributor. Always have serial number and model handy for the appliance you 

are referring to when contacting the distributor.  

For warranty information, Arctiko refers to your distributors terms and 

conditions.

Spare parts

For requirements of spare parts, contact your Arctiko distributor. Please 

inform the serial number of the appliance and model when contacting the 

distributor. 

Arctiko strive for day-to-day delivery of spare parts. However, some special 

parts may take longer due to production time. Arctiko guarantees availability 

of spare parts for all appliances for at least 10 years after the delivery.

Service

Arctiko recommends that service is performed by authorized service

personnel at least once a year. 

Contact your Arctiko distributor for contact information. Always have the  

serial number of the appliance and model ready for the distributor.

Warranty

After-sales

6. Leave the appliance open for a minimum of 24 hours allowing the ice to 

    melt. 

7. When the defrosting is completed thoroughly clean the appliance inside. 

   Wipe off all remaining water and leave the door open until the cabinet is 

    completely dry. 

8. Connect the main power cable to the power socket and switch on the 

   appliance on the display.  

9. Reload the appliance after reaching the set temperature (pull down).

Summary of Contents for ULUF P390

Page 1: ...ORIGINAL OPERATING INSTRUCTIONS Integraline RANGE ULUF P390 P500 P610 P820...

Page 2: ...vents Compressor failure alarm Outcome frames Contents Page 4 4 4 4 5 5 5 6 6 6 6 7 7 8 8 9 10 11 12 13 15 15 16 16 17 17 18 18 18 19 20 21 22 Warnings How to turn on the device Menu user How to chang...

Page 3: ...ctiko is obligated as a supplier to ensure the users safety when operating one of our appliances To prevent personal and material injury or damage please follow the instructions in this manual Intende...

Page 4: ...Be careful not to tip over the appliance during transport to prevent damage or injury Plastic bags pose a suffocation risk Keep away from children Inactivity for extended period If the appliance must...

Page 5: ...n or der to achieve optimal operating results Firm and levelled floor Placing the appliance on a firm and flat floor reduces the risk of excessive noise and vibration Away from any heat sources Avoid...

Page 6: ...connection is a 3 prong connection use a three pin plug and connect the conductor with yellow green insulation to ground In order to maintain a stable operation of the unit voltage variation cannot d...

Page 7: ...filter less construc tion for less maintenance low noise level low energy consumption low heat dissipation and excellent temperature stability The compressors and cooling system is located in the top...

Page 8: ...r openings and control the lights 15 Cover for noise reduction and air flow management 5 Probe See Probes and portholes 16 Remote alarm connection For connecton of additional alarm equipment 6 Door ha...

Page 9: ...ve keyboard You can access a variation of settings in order to adjust your unit for your exact needs A 4GB microSD is provided as standard installed internally for recording functional data every 30 s...

Page 10: ...19 18 Info events Door Warning Action Icons Temperature alarms...

Page 11: ...oller The other compressor will keep running To reset the alarm and reenable the compressor accept the alarm and turn the unit off and on again During pulldown this alarm may be activated If this happ...

Page 12: ...23 22 Outcome frames The outcome frames will be accompanied by a long beep Warnings To accept a warning press DOOR OPEN...

Page 13: ...press Enter Menu user STAND BY SETPOINT LIMITS ACCESS LIST ECOMODE EVENTS GRAPHICS DEFROST SETTINGS SERVICE press and hold to turn off the controller change the temperature setpoint set the high low...

Page 14: ...27 26 How to change the setpoint How to change the alarm limits Press the Menu key to access the menu...

Page 15: ...29 28 Access list Ecomode setting KEY FUNCTION...

Page 16: ...30 Event list If the event list is empty the icon above will be shown Shows number of pages with events Press or to change page Graphs SCROLLING THE FRAPH AND ENABLING THE TRACES REPRESENTATION CRITER...

Page 17: ...edit the telephone numbers where to send the alert messages set the language of user interface edit machine name by alphanumeric keyboard How to set date hour Press select to move the highlight and t...

Page 18: ...is 111 Service For accessing the service menu The service settings are reserved for qualified Arctiko service personnel By entering the admin password both the admin and user passwords can be changed...

Page 19: ...Pressing the backup icon shows the usb pen insertion request frame with 20 second time out If the usb pen has been recognised the setting page of the months to download is displayed otherwise the disp...

Page 20: ...stributor 2 To convert the data open the program Thermo Converter To convert the file press the DAT TO CSV CONVERT button 1 Insert the USB drive with the loaded data in a computer 3 Open the file by f...

Page 21: ...ght corner of the compressor department 1 Logo 2 Contact information 3 Model 4 Voltage range 5 Refrigerant and quantity 6 WEEE Waste from Electrical and Electronic Equipment 7 Compressor type 8 CE mar...

Page 22: ...operation 1h RELAIS U2 U2 Relay status COMP 24h Compressor operation 24h RELAIS U3 U3 Relay status COMP ON Time on compressor RELAIS U4 U4 Relay status COMP OFF Time o compressor RELAIS U5 U5 Relay st...

Page 23: ...ssor compartment using only a dry brush or a vacuum cleaner Parts in the refrigeration system for this appliance are completely sealed These do not require any lubrication If you would like more infor...

Page 24: ...unit away from the sunlight or heat source Frequent door openings may cause the temperature to rise Check if there has been frequent door openings Leave the door closed until the temperature in the un...

Page 25: ...5080063 02 GB Arctiko A S Oddesundcej 39 DK 6715 Esbjerg N Phone 45 7020 0328 info arctiko com arctiko com...

Reviews: