2.3 Environmental verification

Check and verify the following points:

ENVIRONMENT

CHECK (

)

COMMENTS

Ambient condi-

tions

Ensure by using a thermometer a maximum      

ambient temperature of:
+25°C (freezer models)                             

+43°C (refrigerator models)
Ensure no direct sunlight on the appliance
Ensure that the appliance is not located near 

heat sources
Ensure that the room is well ventilated

Surface

Ensure installation on a solid and flat surface 
to avoid vibrations and unintentional noise

Air circulation

Ensure free space around the appliance to 
ensure the required air circulation according 

to the operating instructions manual

CHECKED/TESTED BY

DATE

SIGNATURE

VERIFIED BY

DATE

SIGNATURE

Summary of Contents for LFE 1400

Page 1: ...LRE 700 1400 LFE 700 1400 IQ OQ INSTALLATION AND OPERATIONAL QUALIFICATION...

Page 2: ...intended to verify that the appliance is delivered in corre spondence with the ordered appliance and is installed in a correct environment The Operational Qualification is intended to verify the func...

Page 3: ...e check points 5 2 3 Environmental verification 7 3 Operational Qualification 8 3 1 Parameter verification and adjustment 8 3 2 Placement of test probes 9 3 3 Temperature mapping 11 3 4 Alarm and temp...

Page 4: ...____________________ Date of qualification DD MM YYYY _________________________ Qualification carried out by name _________________________ Customer _________________________ Location of the appliance...

Page 5: ...shall fill out the signature list below in order to establish traceability to the test signatures All devices used for the execution of the current IQ OQ shall be noted in below device list SIGNATURE...

Page 6: ...verify the following points DOCUMENTATION CHECK COMMENTS Ensure operating instruction manual has been supplied with the appliance Ensure operating instruction manual has been read and un derstood by t...

Page 7: ...ped with levelling feet Controller Ensure no scratches on the display Ensure that all cables are mounted according to safety requirements Remote alarm connection Ensure that the remote alarm contact i...

Page 8: ...nsure no direct sunlight on the appliance Ensure that the appliance is not located near heat sources Ensure that the room is well ventilated Surface Ensure installation on a solid and flat surface to...

Page 9: ...E THE VALUE COMMENTS Password PS 22 22 Set point St 4 20 Control delta hysteresis rd 2 2 Interval between defrost dl N A 6 hrs High tempera ture alarm by pass time after defrost and door open d8 0 0 A...

Page 10: ...illustration see the red arrows Each test probe must be loaded with a 25 gr brass block or equivalent to ensure precise temperature measurement LRE LFE 700 PROBE LOCATION Probe no 1 45 cm distance fro...

Page 11: ...Probe no 3 45 cm distance from the bottom of the interior cabinet left side of cabinet Probe no 4 45 cm distance from the top of the interior cabinet right side of cabinet Probe no 5 72 cm distance f...

Page 12: ...on of the appliance by making sure that the temperature of the appliance equals the ambient temperature Start the temperature logging of the test probes Turn on the appliance and let it run until stab...

Page 13: ...he calibration value and repeat above steps Calibration parameter c1 Calibration requires the following field to be signed to confirm calibration has been performed Date Signature Temperature of test...

Page 14: ...of the door open alarm bypass parameter d8d is 2 min Open the door and note start time and start tem perature Leave the door open until the door open alarm activates acoustic alarm and visual error co...

Page 15: ...rature value within the normal operating tempera ture set point hystere sis Temperature of test probe Recovery time Time of door closing time of reaching a temper ature value within the nor mal operat...

Page 16: ...he AL parameter is a relative threshold to the set point see examples in point 3 1 Approval criteria Refrigerator Activation of the acoustic and visu al LO alarm according to the AL low temp alarm Ad...

Page 17: ...ctive action has been initiated and or finalized In case of deviations which cannot be corrected on site please contact the distributor of the appliance for support in correction of the deviation s DE...

Page 18: ...4 Result of Installation and Operational Qualification PRODUCT INFORMATION Manufacturer Model Serial number RESULT CHECK Passed Failed CHECKED TESTED BY DATE SIGNATURE AND STAMP VERIFIED BY DATE SIGN...

Page 19: ...5084132 03...

Reviews: