
9/17
5084132-03
3.2 Placement of test probes
Prepare the appliance for the Operational Qualification by mounting test probes. Place the test
probes inside the appliance as described in below table and as shown in below illustration (see
the red arrows).
Each test probe must be loaded with a 25 gr. brass block
(or equivalent) to ensure precise temperature measurement.
LRE/LFE 700
PROBE
LOCATION
Probe no. 1
45 cm distance from the top of the interior cabinet
Probe no. 2
Center of the chamber (72 cm distance from the top of the interior cabinet)
Probe no. 3
45 cm distance from the bottom of the interior cabinet
Probe no. 1
Probe no. 2
Probe no. 3