3.4 Alarm and temperature recovery verification
The following section verifies the alarm functions and the temperature recovery after door open-
ings.
Before initializing and in between the following tests, allow the appliance to stabilize.
OPEN DOOR ALARM, REMOTE ALARM AND TEMPERATURE RECOVERY TEST
PROCEDURE
RESULT
PASSED (YES/NO)
Verify the function of the
door open alarm
and the
remote alarm contact
by performing the following
steps.
Connect an ohmmeter to the remote alarm contact
terminals 16 and 18 (N/C) placed in the engine room.
Note that the setting of the door open alarm bypass
(parameter d8d) is 2 min.
Open the door and note
start time
and
start tem-
perature
. Leave the door open until the door open
alarm activates (acoustic alarm and visual error code
“dor” on the display).
Verify that the ohmmeter registers and open connec-
tion.
Note the time and display temperature of alarm acti-
vation.
Close the door. The alarm disables automatically.
Approval criteria
-
Open door alarm:
Acoustic and visual alarm
activation time according to the parameter set-
ting of the open door alarm bypass (parameter
d8d)
-
Remote alarm:
open connection is registered
Start time:
Start temperature (appli-
ance):
Set point:
Open signal (Y/N):
Time of alarm
activation:
Display temperature by
alarm activation:
Time difference (min.):
(start time – time of alarm
activation)